Huang X, Liu D M, Liu C, Fu J, Wan B N, Lyu B, Wu Z W, Holcomb C T, Ko J, Rowan W L, Huang H, Miao G Z
School of Electrical Engineering and Automation, Hefei University of Technology, Hefei 230009, China.
Institute of Plasma Physics, Chinese Academy of Sciences, Hefei 230031, China.
Rev Sci Instrum. 2018 Oct;89(10):10B108. doi: 10.1063/1.5036938.
The motional Stark effect (MSE) diagnostic is applied to measure the safety factor q and current density profile of a tokamak device, which are important parameters in realizing the high-performance and long-pulse steady state of a tokamak. A single-channel MSE diagnostic based on dual photoelastic modulators, whose sightline meets with the neutral beam injection at a major radius of R = 2.12 m, has been built for the D window of the Experimental Advanced Superconducting Tokamak (EAST). According to the requirements of MSE diagnostic polarimetric calibration, a high-precision four-dimensional calibration turntable, driven by four stepping motors and controlled by software running on the computer, was designed for EAST. The turntable allows us to rapidly calibrate the MSE diagnostic in a series of positions and angles during EAST maintenance. The turntable can move in four dimensions of translation, yaw, pitch, and roll of the polarizer and can create linearly polarized light at any given angle with accuracy of ∼0.05° for the MSE system offline calibration. The experimental results of the MSE diagnostic calibration in the laboratory show that the turntable has the advantages of high positioning accuracy, flexible spatial movement, and convenient control and fully meets the calibration requirements of an MSE diagnosis system.
运动斯塔克效应(MSE)诊断技术被用于测量托卡马克装置的安全因子q和电流密度分布,这些是实现托卡马克高性能和长脉冲稳态的重要参数。基于双光弹调制器构建了一个单通道MSE诊断系统,其视线在R = 2.12 m的大半径处与中性束注入相交,该系统用于实验先进超导托卡马克(EAST)的D窗口。根据MSE诊断偏振校准的要求,为EAST设计了一个由四个步进电机驱动并由计算机上运行的软件控制的高精度四维校准转台。该转台使我们能够在EAST维护期间在一系列位置和角度快速校准MSE诊断系统。转台可以在偏振器的平移、偏航、俯仰和滚动四个维度上移动,并能以约0.05°的精度在任何给定角度产生线偏振光,用于MSE系统的离线校准。在实验室进行的MSE诊断校准实验结果表明,该转台具有定位精度高、空间运动灵活、控制方便等优点,完全满足MSE诊断系统的校准要求。