It is unknown whether it is possible to measure implant stability with polyetheretherketone (PEEK) healing abutments using electronic percussive testing (EPT).
To investigate the reliability of the percussive test values (PTVs) measured with PEEK healing abutments and to compare them with the PTVs measured with titanium healing abutments.
Thirty dental implants were inserted into the fresh pelvis belonging to a cow. Titanium healing abutments (2 and 5 mm), PEEK healing abutments (5 mm), and prepable standard titanium abutments (5 mm) were screwed to the implants, respectively, and PTVs were measured by two examiners using a wireless EPT device. Differences in PTVs between different dental implant components were evaluated using Friedman's test with post hoc Wilcoxon signed-rank test and Bonferroni correction. Inter and intra-observer reliabilities were detected using interclass correlation coefficients (ICCs) RESULTS: The mean PTVs obtained using the PEEK healing abutments were significantly higher than the mean PTVs obtained using the other abutments for both examiners (P < .01). The ICCs for intra-observer reliability were detected as poor for PEEK healing abutments; and excellent for the other abutment types for both examiners. The ICCs for the inter-observer reliability between the two examiners were poor (0.25) for PEEK healing abutments, and excellent for the 2 and 5 mm titanium healing abutments and for standard abutments (0.82, 0.84 and 0.93, respectively).
Within the limitations of this in vitro study, it may be concluded that EPT measurements should be avoided using PEEK healing abutments because of the poor reliability.
目前尚不清楚是否可以使用电子敲击测试(EPT)测量聚醚醚酮(PEEK)愈合基台的种植体稳定性。
研究使用 PEEK 愈合基台测量的敲击测试值(PTV)的可靠性,并将其与使用钛愈合基台测量的 PTV 进行比较。
将 30 颗牙种植体植入新鲜牛骨盆内。分别将钛愈合基台(2 毫米和 5 毫米)、PEEK 愈合基台(5 毫米)和可预成标准钛基台(5 毫米)拧入种植体上,由两位检查者使用无线 EPT 设备测量 PTV。使用 Friedman 检验和事后 Wilcoxon 符号秩检验以及 Bonferroni 校正评估不同牙种植体部件之间的 PTV 差异。使用组内相关系数(ICC)检测观察者内和观察者间的可靠性。
对于两位检查者,使用 PEEK 愈合基台获得的平均 PTV 均显著高于使用其他基台获得的平均 PTV(P<.01)。对于 PEEK 愈合基台,检查者内可靠性的 ICC 较差;对于其他基台类型,两位检查者的 ICC 均为优秀。两位检查者之间的观察者间可靠性的 ICC 较差(0.25),对于 2 毫米和 5 毫米钛愈合基台和标准基台的 ICC 为优秀(0.82、0.84 和 0.93)。
在本体外研究的限制内,可以得出结论,由于可靠性差,应避免使用 PEEK 愈合基台进行 EPT 测量。