Suppr超能文献

再次探讨使用奥尼克斯胶进行球囊辅助栓塞脑动静脉畸形的安全性和可行性:使用Scepter XC球囊微导管的个人经验

Safety and Feasibility of Balloon-Assisted Embolization with Onyx of Brain Arteriovenous Malformations Revisited: Personal Experience with the Scepter XC Balloon Microcatheter.

作者信息

Jagadeesan Bharathi D, Grande Andrew W, Tummala Ramachandra P

机构信息

Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.

Department of Neurosurgery, University of Minnesota, Minneapolis, Minnesota, USA.

出版信息

Interv Neurol. 2018 Oct;7(6):439-444. doi: 10.1159/000490579. Epub 2018 Jul 13.

Abstract

BACKGROUND/OBJECTIVE: Compliant dual-lumen balloon microcatheters have been used to perform balloon-assisted embolization (BAE) of brain arteriovenous malformations (AVMs) with ethylene vinyl alcohol copolymer (Onyx). However, vessel rupture and microcatheter retention have been reported from BAE using these microcatheters. Using an extra-compliant balloon microcatheter (Scepter XC; Microvention, Tustin, CA, USA) could help avoid pial vessel rupture during BAE. We herein report our experience using this balloon microcatheter for BAE.

METHODS

This retrospective study included patients who underwent BAE of brain AVMs at our institution between June 2012 and March 2017.

RESULTS

The extra-compliant Scepter XC balloon microcatheter was used for BAE of brain AVMs in 23 patients aged 44.3 ± 16.7 years (range 0-65 years). A total of 40 intracranial vessels (39 pial arteries and 1 pial vein) were catheterized and embolized during 30 separate sessions. In all instances, the balloon microcatheter could be successfully advanced to the AVM nidus. A mean volume of 2.4 ± 1.7 mL (range 0.65-4.6 mL) of Onyx was injected per session. There were no instances of vessel rupture, microcatheter retention, or stroke.

CONCLUSION

Utilization of the extra-compliant balloon microcatheter results in safe and effective BAE, which adds to the growing experience with BAE for AVM treatment.

摘要

背景/目的:顺应性双腔球囊微导管已被用于使用乙烯乙烯醇共聚物(Onyx)对脑动静脉畸形(AVM)进行球囊辅助栓塞(BAE)。然而,使用这些微导管进行BAE时曾有血管破裂和微导管滞留的报道。使用超顺应性球囊微导管(Scepter XC;美国加利福尼亚州图斯廷市Microvention公司)可能有助于避免BAE过程中软脑膜血管破裂。我们在此报告使用这种球囊微导管进行BAE的经验。

方法

这项回顾性研究纳入了2012年6月至2017年3月在本机构接受脑AVM的BAE治疗的患者。

结果

超顺应性Scepter XC球囊微导管用于23例年龄为44.3±16.7岁(范围0 - 65岁)的脑AVM的BAE。在30次单独的操作中,总共对40条颅内血管(39条软脑膜动脉和1条软脑膜静脉)进行了插管和栓塞。在所有情况下,球囊微导管均可成功推进至AVM病灶。每次操作平均注入2.4±1.7 mL(范围0.65 - 4.6 mL)的Onyx。没有血管破裂、微导管滞留或中风的情况发生。

结论

使用超顺应性球囊微导管可实现安全有效的BAE,这增加了BAE治疗AVM的经验。

相似文献

引用本文的文献

1
Single Centre Initial Experience with the Scepter Mini Balloon Microcatheter.单中心使用 Scepter Mini 球囊微导管的初步经验。
Interv Neuroradiol. 2024 Jun;30(3):389-395. doi: 10.1177/15910199221128442. Epub 2022 Sep 27.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验