Suppr超能文献

非均匀磁化传递(ihMT)的可重复性:一项重测、多中心研究。

Reproducibility of inhomogeneous magnetization transfer (ihMT): A test-retest, multi-site study.

作者信息

Zhang Lei, Chen Tao, Tian Hongzhe, Xue Hongqiang, Ren Huipeng, Li Li, Fan Qing, Wen Baohong, Ren Zhuanqin

机构信息

Department of Radiology, Baoji Center Hospital, Baoji, Shaanxi, People's Republic of China.

Department of Radiology, Zhengzhou University First Affiliated Hospital, Zhengzhou, Henan, People's Republic of China.

出版信息

Magn Reson Imaging. 2019 Apr;57:243-249. doi: 10.1016/j.mri.2018.11.010. Epub 2018 Nov 19.

Abstract

BACKGROUND

Inhomogeneous magnetization transfer (ihMT) has been reported to feature superior sensitivity and specificity for myelin imaging. However, the reproducibility on ihMT has yet been rarely investigated up to date. The purpose of the present study is to assess the multi-center reproducibility and test-retest variability of ihMT in central nervous system.

METHODS

5 volunteers were recruited and scanned twice on three 3.0 T magnetic resonance imaging (MRI) scanners using ihMT with identical scan parameters. The maps of quantitative ihMT (qihMT) and ihMT ratio (ihMTR) for each scan were calculated. Voxel based analysis then was performed to generate qihMT and ihMTR values for major white matter fibers. The intra- and inter-scanner reliability and reproducibility was assessed with intraclass correlation coefficients (ICCs). Bland-Altman method was used to show the level of agreement between two measurement types. Paired t-test and one-way ANOVA test were also used to compare the difference between inter- and intra-scanner, respectively.

RESULTS

In the 10 major white matter tracts areas, the ICCs indicated high intra- and inter-scanner measurement reliability and reproducibility. The Bland-Altman plots together with 95% confidence interval (CI) across all ROIs in the five volunteers also demonstrated good repeatability. No significant inter- and intra-scanner differences were found in Paired t-test and one-way ANOVA tests.

CONCLUSION

Good inter- and intra-scanner reliability and reproducibility of ihMT measurements were observed in this study. These findings support the use of ihMT measurements as biomarkers in multicenter and/or longitudinal studies.

摘要

背景

据报道,非均匀磁化传递(ihMT)在髓鞘成像方面具有更高的敏感性和特异性。然而,迄今为止,关于ihMT的可重复性很少被研究。本研究的目的是评估中枢神经系统中ihMT的多中心可重复性和重测变异性。

方法

招募了5名志愿者,使用相同的扫描参数,在三台3.0T磁共振成像(MRI)扫描仪上对他们进行了两次ihMT扫描。计算每次扫描的定量ihMT(qihMT)和ihMT比率(ihMTR)图。然后进行基于体素的分析,以生成主要白质纤维的qihMT和ihMTR值。使用组内相关系数(ICC)评估扫描仪内和扫描仪间的可靠性和可重复性。采用Bland-Altman方法显示两种测量类型之间的一致性水平。还分别使用配对t检验和单因素方差分析来比较扫描仪间和扫描仪内的差异。

结果

在10个主要白质束区域,ICC表明扫描仪内和扫描仪间测量具有较高的可靠性和可重复性。5名志愿者所有感兴趣区域的Bland-Altman图以及95%置信区间(CI)也显示出良好的可重复性。配对t检验和单因素方差分析未发现扫描仪间和扫描仪内存在显著差异。

结论

本研究观察到ihMT测量在扫描仪间和扫描仪内具有良好的可靠性和可重复性。这些发现支持在多中心和/或纵向研究中使用ihMT测量作为生物标志物。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验