Succari M, Garric B, Pontezière C, Meneguzzer E, Miocque M
Laboratoire de Biochimie, Hôpital Corentin Celton, Issy-Les-Moulineaux, France.
Pathol Biol (Paris). 1988 Jun;36(6):854-8.
A diazoreaction method of estimation of bilirubin has been adapted on a FP9 parallel analyzer. Biotrol's kit could be used only after modifications of the protocol. The critical study of this adaptation included trials of within run precision (CV 1.93 to 4.42%), total precision (CV 2.59 to 6.66%) as also evaluation of analytical range: the limit of linearity is 340 mumol/l. The evaluation of inaccuracy performed with patient specimens leads to establishment of follow-up norms and interpretation norms of allometry line. Our whole results are in conformity with the performance standards of the protocol for the validation of methods published by the Société Française de Biologie Clinique. Finally the described method is quick acting, reliable and very inexpensive.
一种胆红素测定的重氮反应方法已在FP9平行分析仪上适配。只有在修改方案后才能使用Biotrol试剂盒。对这种适配的关键研究包括批内精密度试验(变异系数为1.93%至4.42%)、总精密度试验(变异系数为2.59%至6.66%)以及分析范围评估:线性极限为340μmol/L。使用患者样本进行的不精密度评估得出了异速生长线的随访规范和解释规范。我们的整体结果符合法国临床生物学会发布的方法验证方案的性能标准。最后,所描述的方法起效快、可靠且成本极低。