三种附着系统在脱位循环后的固位力和磨损特性

Retention Force and Wear Characteristics of three Attachment Systems after Dislodging Cycles.

作者信息

Marin Danny Omar Mendoza, Leite Andressa Rosa Perin, Oliveira Junior Norberto Martins de, Paleari André Gustavo, Pero Ana Carolina, Compagnoni Marco Antonio

机构信息

Department of Dental Materials and Prosthodontics, UNESP - Univ Estadual Paulista, Araraquara, SP, Brazil.

Department of Restorative Dentistry, UNIFAL - Universidade Federal de Alfenas, Alfenas, MG, Brazil.

出版信息

Braz Dent J. 2018 Nov-Dec;29(6):576-582. doi: 10.1590/0103-6440201802074.

Abstract

Several attachment systems for mandibular implant-supported overdentures are currently available and studies are required to understand their mechanical properties. The objective of this study was to evaluate the retention force and wear characteristics of three attachment systems in a simulation of the cyclic dislodging of implant-supported overdentures. Thirty samples were fabricated and divided into 3 groups: 1-O-ring; 2-Mini Ball; and 3-Equator. A mechanical fatigue test was applied to the specimens using a servo-hydraulic universal testing machine performing 5500 insertion/removal cycles (f=0.8 Hz), immersed in artificial saliva. Retention force values ​​were obtained before and after 1500, 3000, and 5500 cycles using a speed of 1 mm/min and a load cell of 1 kN. One specimen from each group was randomly selected and analyzed by scanning electron microscopy. Two-way repeated measures ANOVA and the Bonferroni post hoc test were used for statistical analyses (α=0.05). The O-ring system remained stable during all periods tested and exhibited significantly lower retention force values than the Mini Ball and Equator systems. The Mini Ball system exhibited a significant increase in retention force after the mechanical test (baseline=21.04±3.29N; 5500 cycles=24.01±3.30N).The Equator system exhibited a significant decrease in retention force after each period tested, but the values were higher than the other systems. The type of attachment was found to influence retention force in different ways after mechanical tests. The Equator system exhibited the highest retention force values. The Mini Ball and Equator matrices produced deformation and wear on the surfaces without breakage of the polyamide rings.

摘要

目前有几种用于下颌种植体支持覆盖义齿的附着系统,需要开展研究以了解其力学性能。本研究的目的是在模拟种植体支持覆盖义齿的周期性脱位过程中,评估三种附着系统的固位力和磨损特性。制作了30个样本并分为3组:1 - O形环组;2 - 迷你球组;3 - 赤道组。使用伺服液压万能试验机对样本进行机械疲劳测试,进行5500次插入/取出循环(f = 0.8Hz),样本浸泡在人工唾液中。在1500、3000和5500次循环前后,以1mm/min的速度和1kN的测力传感器获取固位力值。从每组中随机选取一个样本,通过扫描电子显微镜进行分析。采用双向重复测量方差分析和Bonferroni事后检验进行统计分析(α = 0.05)。O形环系统在所有测试期间均保持稳定,并且其固位力值显著低于迷你球组和赤道组系统。迷你球系统在机械测试后固位力显著增加(基线 = 21.04±3.29N;5500次循环 = 24.01±3.30N)。赤道组系统在每个测试周期后固位力均显著下降,但其值高于其他系统。结果发现,附着类型在机械测试后以不同方式影响固位力。赤道组系统表现出最高的固位力值。迷你球组和赤道组的基质表面产生了变形和磨损,但聚酰胺环未破裂。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验