协议优化下铋屏蔽在 CT 胸部检查中的效果。

The effectiveness of bismuth breast shielding with protocol optimization in CT Thorax examination.

机构信息

Department of Physics, Faculty of Science, Universiti Putra Malaysia, 43400 Serdang, Selangor, Malaysia.

Department of Radiology, National Cancer Institute, Precint 7, 62250 Putrajaya, Malaysia.

出版信息

J Xray Sci Technol. 2019;27(1):139-147. doi: 10.3233/XST-180397.

Abstract

BACKGROUND

Numerous techniques had been proposed to reduce radiation exposure in computed tomography (CT) including the use of radiation shielding.

OBJECTIVE

This study aims to evaluate efficacy of using a bismuth breast shield and optimized scanning parameter to reduce breast absorbed doses from CT thorax examination.

METHODS

Five protocols comprising the standard CT thorax clinical protocol (CP1) and four modified protocols (CP2 to CP5) were applied in anthropomorphic phantom scans. The phantom was configured as a female by placing a breast component on the chest. The breast component was divided into four quadrants, where 2 thermoluminescence dosimeters (TLD-100) were inserted into each quadrant to measure the absorbed dose. The bismuth shield was placed over the breast component during CP4 and CP5 scans.

RESULTS

The pattern of absorbed doses in each breast and quadrant were approximately the same for all protocols, where the 4th quadrant > 3rd quadrant > 2nd quadrant > 1st quadrant. The mean absorbed dose value in CP3 was reduced to almost 34% of CP1's mean absorbed dose. It was reduced even lower to 15% of CP1's mean absorbed dose when the breast shield was used in CP5.

CONCLUSION

This study showed that CT radiation exposure on the breast could be reduced by using a bismuth shield and low tube potential protocol without compromising the image quality.

摘要

背景

为了降低计算机断层扫描(CT)中的辐射暴露,已经提出了许多技术,包括使用辐射屏蔽。

目的

本研究旨在评估使用铋乳腺屏蔽和优化扫描参数来降低 CT 胸部检查中乳腺吸收剂量的效果。

方法

在人体模型扫描中应用了五个方案,包括标准 CT 胸部临床方案(CP1)和四个修改方案(CP2 至 CP5)。通过在胸部放置乳腺组件来将模型配置为女性。乳腺组件被分为四个象限,每个象限插入两个热释光剂量计(TLD-100)以测量吸收剂量。在 CP4 和 CP5 扫描期间,将铋屏蔽置于乳腺组件上方。

结果

对于所有方案,每个乳房和象限的吸收剂量模式大致相同,其中第四象限>第三象限>第二象限>第一象限。CP3 的平均吸收剂量值降低到 CP1 的平均吸收剂量的近 34%。当在 CP5 中使用乳腺屏蔽时,它甚至降低到 CP1 的平均吸收剂量的 15%。

结论

本研究表明,通过使用铋屏蔽和低管电压方案,可以降低乳腺 CT 辐射暴露,而不会影响图像质量。