RADPAD防护巾在减少导管室辐射暴露方面的效果——印度的首项研究

Efficacy of RADPAD protection drape in reducing radiation exposure in the catheterization laboratory-First Indian study.

作者信息

Shah Pujan, Khanna Roopali, Kapoor Aditya, Goel Pravin K

机构信息

Sanjay Gandhi PGIMS, Lucknow, India.

Sanjay Gandhi PGIMS, Lucknow, India.

出版信息

Indian Heart J. 2018 Dec;70 Suppl 3(Suppl 3):S265-S268. doi: 10.1016/j.ihj.2018.03.008. Epub 2018 Mar 27.

Abstract

BACKGROUND

Occupational radiation exposureis a growing problem due to increasing number and complexity of interventional procedures.The RADPAD is a lead-free sterile drape containing bismuth and barium that reduces scatter radiation during fluoroscopic procedures. We aimed to study the radiation exposure reduction to operators with the use of RADPAD and also measureradiation doses in different angiographic projections.

METHODS

65 randomly selected patients undergoing elective complex percutaneous coronary intervention (PCI) procedures from January 2017 to 2017 were randomized in a 1:1 pattern with or without the RADPAD. Primary endpoint was the ratio of operator received dose in mrem to total radiation in Gyat the end of the procedure which was designated ''Relative operator exposure'', with or without RADPAD.

RESULTS

Despite similar fluoroscopy times (20.4 ± 9.4 min with RADPAD vs. 19.4 ± 9.2 min without RADPAD, P = 0.871) and total radiation dose (3.4 ± 4.3 Gy with RADPAD vs. 2.3 ± 1.4 Gy, P = 0.198), the relative operator exposure was significantly less with RADPAD (1.39 ± 0.95) as compared to no RADPAD group (2.27 ± 1.4) (p = 0.004) amounting to a 39% reduction. Additionally mean radiation dose per shoot of recorded Left anterior oblique (LAO) oriented projections was 34.4 ± 15.7mGyvs 24.9 ± 12.9 mGy for a non LAO oriented projection. (p < 0.001).

CONCLUSION

RADPAD significantly reduces radiation exposure to the primary operator during prolonged complex PCI procedures. Further, amongst all views, LAO views have significantly higher emitted radiation as compared to Non LAO views and need more radiation protection.

摘要

背景

由于介入手术数量的增加和复杂性,职业辐射暴露成为一个日益严重的问题。RADPAD是一种不含铅的无菌手术单,含有铋和钡,可减少透视检查过程中的散射辐射。我们旨在研究使用RADPAD对操作人员辐射暴露的降低情况,并测量不同血管造影投影中的辐射剂量。

方法

从2017年1月至2017年随机选择65例行择期复杂经皮冠状动脉介入治疗(PCI)手术的患者,按1:1的比例随机分为使用或不使用RADPAD两组。主要终点是手术结束时操作人员接受的剂量(以毫雷姆为单位)与总辐射剂量(以戈瑞为单位)的比值,即“相对操作人员暴露”,分为使用或不使用RADPAD两种情况。

结果

尽管透视时间相似(使用RADPAD时为20.4±9.4分钟,不使用RADPAD时为19.4±9.2分钟,P = 0.871)和总辐射剂量相似(使用RADPAD时为3.4±4.3戈瑞,不使用RADPAD时为2.3±1.4戈瑞,P = 0.198),但与不使用RADPAD组(2.27±1.4)相比,使用RADPAD时的相对操作人员暴露显著降低(1.39±0.95)(p = 0.004),降低了39%。此外,记录的左前斜位(LAO)投影每次拍摄的平均辐射剂量为34.4±15.7毫戈瑞,而非LAO位投影为24.9±12.9毫戈瑞。(p < 0.001)。

结论

在长时间的复杂PCI手术中,RADPAD可显著降低对主操作人员的辐射暴露。此外,在所有视图中,LAO视图的发射辐射明显高于非LAO视图,需要更多的辐射防护。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/487c/6310698/a0a3d2ceb379/gr1.jpg