University of Amsterdam, Van 't Hoff Institute for Molecular Sciences, Analytical Chemistry-Group, P.O. Box 94157, 1090 GD, Amsterdam, the Netherlands.
University of Amsterdam, Van 't Hoff Institute for Molecular Sciences, Analytical Chemistry-Group, P.O. Box 94157, 1090 GD, Amsterdam, the Netherlands; Unilever Research and Development, P.O. Box 114, 3130 AC Vlaardingen, the Netherlands.
J Chromatogr A. 2019 Apr 12;1590:113-120. doi: 10.1016/j.chroma.2019.01.015. Epub 2019 Jan 8.
Highly purified mineral oils are used in several pharmaceutical, foods and cosmetics applications. A fast and simple method was developed for the analysis of the total level of residual mineral oil aromatic hydrocarbons (MOAH) in these oils and in the intermediate oils that were sampled during the purification process. The method is based on gas chromatography with vacuum ultraviolet detection (GC-VUV) and relies on the spectral differences between the aliphatic and aromatic compounds in the sample. Because the detector provides a good selectivity for aromatics, direct quantification of the MOAH content is possible without the need for a laborious preseparation of the mineral oil. The method was successfully applied for the direct analysis of the MOAH levels of 18 different mineral oil samples. The aromatics contents obtained by GC-VUV were similar to those obtained using two conventional methods (NPLC-GC-FID and SPE-GC-FID), with no statistically significant difference. The detector response was linear over the concentration range tested (0.5-20 mg/mL) and the repeatability (RSD value) was less than 8%, which is better than the typical values for the conventional methods (up to 15% RSD). The minimum MOAH level that can be determined with this method is approximately 0.13%, making the GC-VUV method sufficiently sensitive for the analysis of all but the highest purity mineral oils.
高度精制的矿物油在几种医药、食品和化妆品应用中使用。本研究开发了一种快速简单的方法,用于分析这些油和在纯化过程中采样的中间油中残留矿物油芳香烃(MOAH)的总水平。该方法基于气相色谱与真空紫外检测(GC-VUV),依赖于样品中脂肪族和芳香族化合物之间的光谱差异。由于检测器对芳烃具有良好的选择性,因此可以直接定量 MOAH 含量,而无需对矿物油进行繁琐的预分离。该方法成功地应用于直接分析 18 种不同矿物油样品的 MOAH 水平。通过 GC-VUV 获得的芳烃含量与使用两种常规方法(NPLC-GC-FID 和 SPE-GC-FID)获得的结果相似,没有统计学上的显著差异。在测试的浓度范围内(0.5-20mg/mL),检测器响应呈线性,重复性(RSD 值)小于 8%,优于常规方法的典型值(高达 15%RSD)。用这种方法可以检测到的最低 MOAH 水平约为 0.13%,使得 GC-VUV 方法对所有除了最高纯度的矿物油的分析都足够灵敏。