This is a report of a prospective study in which 41 consecutive autologous tissue transplants were monitored using the Laserflo BPM model 403 laser Doppler monitor (TSI Inc., St. Paul, MN). Flaps were monitored both intraoperatively and postoperatively, and clinical flap monitoring was compared with laser Doppler monitoring. Twenty-nine flaps had no problems as indicated by clinical or Doppler measurement. Twelve flaps underwent 13 reexplorations; one flap was explored twice. Five venous and five arterial anastomoses were corrected, with complete survival of the flaps. Three hematomas were evacuated. Two of these flaps failed. There were no instances where the laser Doppler indicated a problem with flap viability that subsequently did not merit reexploration. There were no instances of flap failure undetected by the laser Doppler. In comparison to the clinical monitoring, laser Doppler monitoring was more rapid and more precise in identifying problems with the flap. The use of this monitor has improved our salvage rate for flap reexploration from 50 to 85% and has provided a more objective standard for both nurses and physicians to evaluate flaps in the perioperative period.
这是一份前瞻性研究报告,该研究使用型号为403的Laserflo BPM激光多普勒监测仪(TSI公司,明尼苏达州圣保罗)对41例连续的自体组织移植进行了监测。在术中及术后均对皮瓣进行监测,并将临床皮瓣监测与激光多普勒监测进行比较。29例皮瓣在临床或多普勒测量中均未出现问题。12例皮瓣进行了13次再次探查;1例皮瓣进行了两次探查。纠正了5例静脉和5例动脉吻合口,皮瓣完全存活。清除了3例血肿。其中2例皮瓣失败。没有出现激光多普勒显示皮瓣存活有问题但随后不值得再次探查的情况。也没有激光多普勒未检测到的皮瓣失败情况。与临床监测相比,激光多普勒监测在识别皮瓣问题方面更快、更精确。使用该监测仪将我们皮瓣再次探查的挽救率从50%提高到了85%,并为护士和医生在围手术期评估皮瓣提供了更客观的标准。