Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
World Neurosurg. 2019 May;125:e348-e352. doi: 10.1016/j.wneu.2019.01.076. Epub 2019 Jan 28.
To evaluate the safety and accuracy of use of a 3-dimensional printed navigation template in the placement of a cortical bone trajectory (CBT) screw in the middle-upper thoracic spine.
Ten human cadavers were included in the study. Sixty CBT screws were placed on 1 side, using the free-hand technique, and 60 CBT screws were placed on the other side, using the navigation template that was designed and printed using data from 10 cadavers. The safety and accuracy of use of the CBT screws were directly evaluated by radiography and computed tomography.
Computed tomography revealed that 2 and 3 of 60 screws, placed using the navigation template, were broken in the medial or lateral areas and in the superior or inferior pedicle wall, respectively. Furthermore, 8 screws were broken in the medial or lateral areas and 11 screws were broken in the superior or inferior pedicle wall when the free-hand technique was used. Radiography revealed that 3 screws in zone I, 55 screws in zone II, and 2 screws in zone III were placed using the navigation template. Furthermore, 7 screws in zone I, 45 screws in zone II, and 8 screws in zone III were placed using the free-hand technique.
In this cadaver study, insertion of the CBT screws in the middle-upper thoracic spine with the assistance of the navigation template was safe and convenient.
评估在中上胸椎中使用三维打印导航模板进行皮质骨轨迹(CBT)螺钉放置的安全性和准确性。
本研究纳入了 10 具人体尸体。在一侧使用徒手技术放置了 60 个 CBT 螺钉,在另一侧使用了根据 10 具尸体数据设计和打印的导航模板放置了 60 个 CBT 螺钉。通过 X 线和计算机断层扫描直接评估 CBT 螺钉的安全性和准确性。
计算机断层扫描显示,使用导航模板放置的 60 个螺钉中有 2 个和 3 个分别在内侧或外侧区域以及在上下椎弓根壁处断裂。此外,使用徒手技术时,8 个螺钉在内外侧区域断裂,11 个螺钉在上下椎弓根壁处断裂。X 线显示,使用导航模板在 I 区放置了 3 个螺钉,在 II 区放置了 55 个螺钉,在 III 区放置了 2 个螺钉。此外,在 I 区使用徒手技术放置了 7 个螺钉,在 II 区放置了 45 个螺钉,在 III 区放置了 8 个螺钉。
在这项尸体研究中,使用导航模板辅助中上胸椎 CBT 螺钉的插入是安全且方便的。