Suppr超能文献

一种微型化傅里叶变换静电线性离子阱质谱仪:质量范围和分辨率。

A Miniaturized Fourier Transform Electrostatic Linear Ion Trap Mass Spectrometer: Mass Range and Resolution.

机构信息

Department of Chemistry, Purdue University, 560 Oval Drive, West Lafayette, IN, 47907-2084, USA.

出版信息

J Am Soc Mass Spectrom. 2019 Apr;30(4):588-594. doi: 10.1007/s13361-018-02126-x. Epub 2019 Feb 12.

Abstract

Mass resolution (M/ΔM) was increased by reducing the axial length of a Fourier transform electrostatic linear ion trap (FT-ELIT) mass spectrometer. The increase in mass resolution corresponds directly to increased axial ion frequencies in the FT-ELIT. Increased mass resolution was demonstrated for equivalent transient lengths in a 5.25″ versus 2.625″ ELIT using the isotopes of [bradykinin+2H] and [insulin+5H] as test ions. Both bradykinin and insulin show mass resolution increases of ~ 90% allowing baseline resolution of the [insulin+5H] isotopes after only 300 ms of data acquisition. Relative changes in mass/charge range were explored using mirror switching to trap ions injected axially into the ELIT. When trapping ions using mirror switching, the mass/charge range in a FT-ELIT mass spectrometer for a given switch time is determined by the time required for fast ions to enter and exit the trap after one reflection versus the time it takes for slow ions to enter the trap. By reducing the length of the FT-ELIT mass spectrometer while maintaining a constant distance from the point from which ions are initially accelerated to the entrance mirror, only the low m/z limit is affected for a given mirror switching time. For the two ELIT lengths examined here, the effective mass/charge range at any given switch time is reduced from m/z-8.9m/z for the 5.25″ ELIT to m/z-5.2m/z for the 2.625″ ELIT. Graphical Abstract.

摘要

质量分辨率(M/ΔM)通过减小傅里叶变换静电线性离子阱(FT-ELIT)质谱仪的轴向长度来提高。质量分辨率的提高与 FT-ELIT 中轴向离子频率的增加直接对应。在使用 [bradykinin+2H] 和 [insulin+5H] 同位素作为测试离子的情况下,在 5.25 英寸与 2.625 英寸 ELIT 中,通过等效的瞬态长度证明了质量分辨率的提高。Bradykinin 和 insulin 的质量分辨率都提高了约 90%,仅在 300 ms 的数据采集后就可以基线分辨 [insulin+5H] 同位素。使用镜像切换来捕获轴向注入 ELIT 的离子,研究了质量/电荷范围的相对变化。使用镜像切换捕获离子时,FT-ELIT 质谱仪中给定开关时间的质量/电荷范围取决于快速离子在经过一次反射后进入和离开陷阱所需的时间与慢速离子进入陷阱所需的时间之比。通过在保持离子最初加速到入口镜的点的距离不变的情况下减小 FT-ELIT 质谱仪的长度,对于给定的镜像切换时间,只有低 m/z 限制受到影响。对于这里检查的两个 ELIT 长度,在任何给定的切换时间,有效质量/电荷范围从 5.25 英寸 ELIT 的 m/z-8.9m/z 减小到 2.625 英寸 ELIT 的 m/z-5.2m/z。图表摘要。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验