强迫振荡和脉冲振荡测量的比较:临床人群和打印气道模型研究。

Comparison of Forced and Impulse Oscillometry Measurements: A Clinical Population and Printed Airway Model Study.

机构信息

NIHR Biomedical Research Centre: Respiratory Theme and Department of Respiratory Sciences, Universityof Leicester, Leicester, United Kingdom.

School of Physics and Astronomy, University of Nottingham, Nottingham, United Kingdom.

出版信息

Sci Rep. 2019 Feb 14;9(1):2130. doi: 10.1038/s41598-019-38513-x.

Abstract

The use of commercialised forced oscillation (FOT) devices to assess impedance in obstructive diseases such as asthma has gained popularity. However, it has yet to be fully established whether resistance and reactance measurements are comparable across different FOT devices, particularly in disease. We compared two commercially available FOT devices: Impulse Oscillometry (IOS) and TremoFlo FOT (Thorasys) in a) clinical adult population of healthy controls (n = 14), asymptomatic smokers (n = 17) and individuals with asthma (n = 73) and b) a 3D printed CT-derived airway tree model resistance, as well as a 3 L standardised volume reactance. Bland-Altman Plots and linear regressions were used to evaluate bias between the devices. Resistance measurements at both 5 and 20 Hz were numerically higher with IOS compared to FOT, with evidence of small and statistically significant proportional systematic bias and a positive Bland-Altman regression slope at both 5 and 20 Hz. In contrast, the IOS device recorded reactances that were less negative at both 5 Hz and 20 Hz and significantly smaller reactance areas when compared to TremoFlo. Larger statistically significant proportional systematic biases were demonstrated with both reactance at 5 Hz and reactance area (AX) between the devices with a negative Bland-Altman regression slope. The printed airway resistance and standardised volume reactance confirmed the observations seen in patients. We have demonstrated that the impulse oscillation system and TremoFlo FOT demonstrate comparative bias, particularly when comparing airway reactance in patients. Our results highlight the need for further standardisation across FOT measurement devices, specifically using variable test loads for reactance standardisation.

摘要

商业性强迫振荡(FOT)设备在评估哮喘等阻塞性疾病的阻抗方面的应用越来越受欢迎。然而,不同 FOT 设备之间的阻力和电抗测量是否具有可比性尚未得到充分证实,特别是在疾病中。我们比较了两种商业上可用的 FOT 设备:脉冲振荡仪(IOS)和 Tremoflo FOT(Thorasy),a)在临床成年健康对照组(n=14)、无症状吸烟者(n=17)和哮喘患者(n=73)中,b)在 3D 打印 CT 衍生气道树模型阻力以及 3L 标准化体积电抗中。Bland-Altman 图和线性回归用于评估设备之间的偏差。与 FOT 相比,IOS 在 5Hz 和 20Hz 时的阻力测量值更高,存在小而具有统计学意义的比例系统偏差的证据,并且在 5Hz 和 20Hz 时的 Bland-Altman 回归斜率均为正。相比之下,IOS 设备在 5Hz 和 20Hz 时记录的电抗值更小,与 Tremoflo 相比,电抗区域明显更小。在两种情况下,5Hz 时的电抗和电抗面积(AX)之间的设备反应都表现出更大的统计学意义上的比例系统偏差,Bland-Altman 回归斜率为负。打印气道阻力和标准化体积电抗证实了在患者中观察到的结果。我们已经证明,脉冲振荡系统和 Tremoflo FOT 具有相似的偏差,特别是在比较患者气道电抗时。我们的结果强调了需要在 FOT 测量设备之间进行进一步的标准化,特别是使用可变测试负荷进行电抗标准化。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/ac39/6376033/95329a373b01/41598_2019_38513_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验