用于脊柱远端腹部肿瘤的X-sight脊柱追踪系统的性能评估:二维剂量分析

Performance evaluation of the X-sight spine tracking system for abdominal tumors distal to spine: A 2D dosimetric analysis.

作者信息

Eken Selami, Zorlu Faruk, Yeginer Mete, Ozyigit Gokhan

机构信息

Hacettepe University Faculty of Medicine, Department of Radiation Oncology, Ankara 06100, Turkey.

Hacettepe University Faculty of Medicine, Department of Radiation Oncology, Ankara 06100, Turkey.

出版信息

Med Dosim. 2019;44(4):370-374. doi: 10.1016/j.meddos.2019.01.003. Epub 2019 Feb 22.

Abstract

The X-sight spine tracking system was integrated with Cyberknife unit to deliver radiosurgery treatments for spinal tumors without fiducial placement. However the tracking system can also be used for the treatment of abdominal tumors located in a certain distance from the spine. The aim of our study is to evaluate the tracking performance of the X-sight spine system for abdominal tumors distal to spine based on the 3 factors: tumor distance from the reference vertebra, the angle of tumor with the vertebra, and the amplitude of tumor motion due to respiration. An experimental setup was designed mainly with ovine lumbar vertebrae and the BrainLab ExacTrac gating phantom. Planning Target Volume (PTV) structures were delineated at different vertical distances from the reference vertebra. The dosimetric measurements were taken with GafChromic EBT3 film placed between slab phantoms so that they corresponded to centers of the target volumes. Dosimetric comparisons were performed based on dose-volume parameters and the gamma analysis. The measurements were then repeated for the same experimental conditions by using the Synchrony system to compare tracking performances. Using the X-sight system, percentage differences between the dose-volume parameters of the Treatment Planning System (TPS) calculations and the EBT3 film readings went up to 12% for the motion amplitude of 8 mm. The differences decreased with small motions while angles and vertical distances of the lesion locations did not induce major changes in dose discrepancies. Percentages of pixels passing gamma analysis were found to be below the acceptance threshold of 95%. Using the Synchrony system, the measured dose distributions had more similar patterns with those of the TPS system such that the percentage differences in the dose parameters were less than 4% and the gamma passing rates were found to be higher than 95%. Our results showed that the X-sight spine system should not be chosen for tracking abdominal tumors distal to the spine or osseous structures because of the effect of diaphragmatic motion on entire abdominal region. The fiducial-based Synchrony tracking system can be preferred for these tumors.

摘要

X-sight脊柱追踪系统与射波刀设备集成,用于在不放置基准标记的情况下对脊柱肿瘤进行放射外科治疗。然而,该追踪系统也可用于治疗距离脊柱一定距离的腹部肿瘤。我们研究的目的是基于以下三个因素评估X-sight脊柱系统对脊柱远端腹部肿瘤的追踪性能:肿瘤与参考椎体的距离、肿瘤与椎体的角度以及呼吸引起的肿瘤运动幅度。主要采用绵羊腰椎和BrainLab ExacTrac门控体模设计了一个实验装置。在距参考椎体不同垂直距离处勾勒出计划靶区(PTV)结构。使用放置在平板体模之间的GafChromic EBT3胶片进行剂量测量,使其对应于靶区中心。基于剂量体积参数和伽马分析进行剂量学比较。然后在相同实验条件下使用同步系统重复测量,以比较追踪性能。使用X-sight系统时,对于8mm的运动幅度,治疗计划系统(TPS)计算的剂量体积参数与EBT3胶片读数之间的百分比差异高达12%。小幅度运动时差异减小,而病变位置的角度和垂直距离并未引起剂量差异的重大变化。发现通过伽马分析的像素百分比低于95%的接受阈值。使用同步系统时,测量的剂量分布与TPS系统的剂量分布模式更相似,剂量参数的百分比差异小于4%,伽马通过率高于95%。我们的结果表明,由于膈肌运动对整个腹部区域的影响,不应选择X-sight脊柱系统来追踪脊柱或骨性结构远端的腹部肿瘤。对于这些肿瘤,基于基准标记的同步追踪系统可能更合适。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验