CyberKnife Xsight 与基于基准点的靶区追踪系统:动态体模中一种新的 3D 剂量学比较。

CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom.

机构信息

School of Physics, Mathematics and Computing, The University of Western Australia, 35 Stirling Highway, Mailbag M013, Crawley, WA, 6009, Australia.

Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, Australia.

出版信息

Radiat Oncol. 2022 Sep 8;17(1):154. doi: 10.1186/s13014-022-02123-1.

Abstract

BACKGROUND

The CyberKnife Xsight lung-tracking system (XLTS) provides an alternative to fiducial-based target-tracking systems (FTTS) for non-small-cell lung cancer (NSCLC) patients without invasive fiducial insertion procedures. This study provides a method for 3D independent dosimetric verification of the accuracy of the FTTS compared to the XLTS without relying on log-files generated by the CyberKnife system.

METHODS

A respiratory motion trace was taken from a 4D-CT of a real lung cancer patient and applied to a modified QUASAR™ respiratory motion phantom. A novel approach to 3D dosimetry was developed using Gafchromic EBT3 film, allowing the 3D dose distribution delivered to the moving phantom to be reconstructed. Treatments were planned using the recommended margins for one and three fiducial markers and XLTS 2-view, 1-view and 0-view target-tracking modalities. The dose delivery accuracy was analysed by comparing the reconstructed dose distributions to the planned dose distributions using gamma index analysis.

RESULTS

For the 3%/2 mm gamma criterion, gamma passing rates up to 99.37% were observed for the static deliveries. The 3-fiducial and 1-fiducial-based deliveries exhibited passing rates of 93.74% and 97.82%, respectively, in the absence of target rotation. When target rotation was considered, the passing rate for 1-fiducial tracking degraded to 91.24%. The passing rates observed for XLTS 2-view, 1-view and 0-view target-tracking were 92.78%, 96.22% and 76.08%, respectively.

CONCLUSIONS

Except for the XLTS 0-view, the dosimetric accuracy of the XLTS was comparable to the FTTS under equivalent treatment conditions. This study gives us further confidence in the CyberKnife XLTS and FTTS systems.

摘要

背景

CyberKnife Xsight 肺部跟踪系统(XLTS)为非小细胞肺癌(NSCLC)患者提供了一种替代基于标记物的目标跟踪系统(FTTS)的方法,无需进行侵入性标记物插入程序。本研究提供了一种方法,用于在不依赖 CyberKnife 系统生成的日志文件的情况下,对 FTTS 与 XLTS 的准确性进行 3D 独立剂量学验证。

方法

从真实肺癌患者的 4D-CT 中获取呼吸运动轨迹,并将其应用于改良的 QUASAR™呼吸运动体模。使用 Gafchromic EBT3 胶片开发了一种新的 3D 剂量学方法,允许重建输送到移动体模的 3D 剂量分布。使用建议的一个和三个标记物以及 XLTS 2 视图、1 视图和 0 视图目标跟踪模式为治疗计划。使用伽马指数分析将重建的剂量分布与计划的剂量分布进行比较,分析剂量输送的准确性。

结果

对于 3%/2mm 伽马标准,静态输送的伽马通过率高达 99.37%。在没有目标旋转的情况下,基于 3 个标记物和 1 个标记物的输送的通过率分别为 93.74%和 97.82%。当考虑目标旋转时,1 个标记物跟踪的通过率降低至 91.24%。XLTS 2 视图、1 视图和 0 视图目标跟踪的通过率分别为 92.78%、96.22%和 76.08%。

结论

除了 XLTS 0 视图之外,在等效治疗条件下,XLTS 的剂量学准确性与 FTTS 相当。本研究使我们对 CyberKnife XLTS 和 FTTS 系统更有信心。