Sarkisov G S, McChesney P D, Stein S, Pronko S, Bluem H, Lednum G
Raytheon, Missile Systems, Directed Energy Division, 1300 Eubank Blvd, Albuquerque, New Mexico 87123, USA.
Raytheon, Integrated Defense Systems, 50 Apple Hill Drive, Tewksbury, Massachusetts 01876, USA.
Rev Sci Instrum. 2019 Feb;90(2):023504. doi: 10.1063/1.5081830.
To increase the sensitivity of optical interferometry for diagnostic of gas/plasma jet density with weak refractivity, we realize several multi-pass probing setups with the visualization field. These optical setups are based on a modified Michelson interferometer with multi-pass and multi-angle propagation of the probing beam through the transparent inhomogeneity. In comparison to the standard single-pass Mach-Zehnder setup, the multi-pass interferometer allows for an increase in the measurement sensitivity of up to 8 times. This technique can be applied to a transparent axially symmetrical gas/plasma jet with low refractivity.
为了提高光学干涉测量法对弱折射性气体/等离子体射流密度诊断的灵敏度,我们实现了几种带有可视化场的多程探测装置。这些光学装置基于一种经过改进的迈克尔逊干涉仪,探测光束通过透明不均匀体进行多程和多角度传播。与标准的单程马赫曾德尔装置相比,多程干涉仪可将测量灵敏度提高多达8倍。该技术可应用于具有低折射性的透明轴对称气体/等离子体射流。