评估新型六触点骶神经调节系统在猪体内的效果。

Evaluation of sacral neuromodulation system with new six-contact points electrode in pigs.

机构信息

Department of Urology of Capital Medical University, China Rehabilitation Research Center, Capital Medical University, Beijing, China.

Beijing Key Laboratory of Neural Injury and Rehabilitation, Beijing, China.

出版信息

Neurourol Urodyn. 2019 Apr;38(4):1038-1043. doi: 10.1002/nau.23956. Epub 2019 Mar 7.

Abstract

AIMS

We investigated the effects of sacral neuromodulation using the new six-contact electrode vs the four-contact electrode in pigs.

METHODS

Randomly, a six-contact electrode was implanted in eight pigs in one side of the third sacral (S3) foramen, and a four-contact electrode was implanted in the other side using the same method. Using an external neurostimulator, the number of contact points (sensitive voltage ≤ 2 V) of both electrodes (SacralStim and InterStim systems) was calculated. Cystometry was performed by infusing normal saline or acetic acid. Then sacral neuromodulation with the SacralStim and InterStim systems was induced at a voltage at which we could observe perianal and/or tail movement. Multiple cystometrograms were performed to determine the effects of the two systems on the micturition reflex.

RESULTS

The mean number of sensitive points of six-contact electrodes of the SacralStim system (2.63 ± 0.32) was higher than that of the quadripolar-lead electrodes of the InterStim system (1.38 ± 0.18), and the difference was statistically significant (P < 0.05). Acetic acid-induced bladder overactivity significantly reduced bladder capacity to 54.89% ± 4.7% of the normal saline control level. During acetic acid infusion, sacral neuromodulation with the SacralStim system suppressed bladder overactivity and significantly increased bladder capacity to 70.41% ± 5.4% of the normal saline control level, compared with the acetic acid level ( P < 0.05). Moreover, sacral neuromodulation with the InterStim system also significantly increased bladder capacity to 69.63% ± 5.3% of the normal saline control level, compared with the acetic acid level ( P < 0.05). No significant differences were found in the results obtained using the two systems ( P > 0.05).

CONCLUSIONS

The six-contact electrode of the SacralStim system had more sensitive points (<2 V) than that of the quadripolar-lead electrode of InterStim system. Potentially, it has more postimplantation programming options and battery savings manifested by lower voltage will increase the longevity of the stimulator. Further studies of sacral neuromodulation with six-contact electrodes in clinical practice are needed.

摘要

目的

我们研究了使用新型六触点电极和四触点电极对猪进行骶神经调节的效果。

方法

随机选择 8 头猪,一侧第三骶骨(S3)孔植入六触点电极,另一侧使用相同方法植入四触点电极。使用外部神经刺激器计算两个电极(SacralStim 和 InterStim 系统)的触点数(敏感电压≤2V)。通过输注生理盐水或乙酸进行膀胱测压。然后,在我们可以观察到肛门和/或尾巴运动的电压下,使用 SacralStim 和 InterStim 系统进行骶神经调节。进行多次膀胱测压以确定两个系统对排尿反射的影响。

结果

SacralStim 系统六触点电极的平均敏感点数(2.63±0.32)高于 InterStim 系统四极导联电极的(1.38±0.18),差异有统计学意义(P<0.05)。乙酸诱导的膀胱过度活动显著将膀胱容量降低至生理盐水对照水平的 54.89%±4.7%。在乙酸输注期间,与乙酸水平相比,SacralStim 系统的骶神经调节抑制了膀胱过度活动,并将膀胱容量显著增加至生理盐水对照水平的 70.41%±5.4%(P<0.05)。此外,与乙酸水平相比,InterStim 系统的骶神经调节也将膀胱容量显著增加至生理盐水对照水平的 69.63%±5.3%(P<0.05)。两种系统的结果无显著差异(P>0.05)。

结论

SacralStim 系统的六触点电极比 InterStim 系统的四极导联电极具有更多的敏感点(<2V)。植入后可能具有更多的程控选择,较低的电压可节省电池,从而延长刺激器的使用寿命。需要进一步研究六触点电极在临床实践中的骶神经调节。