Magn Reson Med. 2019 Jun;81(6):3691-3704. doi: 10.1002/mrm.27691. Epub 2019 Mar 7.
The goal of this work is to demonstrate 1 mm through-plane resolution in multislice T2SE MRI using k -space processing of overlapping slices and show applicability in prostate MRI.
Multiple overlapped slices are acquired and Fourier transformed in the slice-select direction. The slice profile is incorporated into a Tikhonov-regularized reconstruction. Through-plane resolution is tested in a resolution phantom. An anthropomorphic prostate phantom is used to study the SNR, and results are compared with theoretical prediction. The proposed method is tested in 16 patients indicated for clinical prostate MRI who gave written informed consent as overseen by our IRB. The "proposed" vs. "reference" multislice images are compared using multiple evaluation criteria for through-plane resolution.
The modulation transfer function (MTF) plots of the resolution phantom show good modulation at frequency 0.5 lp/mm, demonstrating 1 mm through-plane resolution restoration. The SNR measurements experimentally match the theoretically predicted values. The radiological evaluation shows that the proposed method is superior to the reference method for five criteria of sharpness but inferior with respect to artifacts.
In conjunction with overlapped slices a k -space-based reconstruction approach can be used to improve through-plane resolution in multislice T2SE MRI. 1 mm resolution is demonstrated from 3.2 mm thick slices. The in vivo results from prostate MRI show improved sharpness when compared to the standard multislice method.
本研究旨在通过重叠切片的 k 空间处理,在多回波 T2SE MRI 中实现 1 毫米的层间分辨率,并展示其在前列腺 MRI 中的应用。
采集多个重叠切片,并在层面选择方向上进行傅里叶变换。将切片轮廓纳入泰克诺诺夫正则化重建中。在分辨率体模中测试层间分辨率。使用仿体前列腺体模研究 SNR,并与理论预测进行比较。该方法在 16 名经临床前列腺 MRI 检查的患者中进行了测试,患者均签署了书面知情同意书,并经我们的 IRB 监督。通过多项层间分辨率评估标准比较“提出的”与“参考的”多回波图像。
分辨率体模的调制传递函数(MTF)图在频率为 0.5 lp/mm 时显示出良好的调制,证明了 1 毫米的层间分辨率恢复。SNR 测量值与理论预测值吻合良好。放射学评估显示,该方法在五个锐利度标准方面优于参考方法,但在伪影方面较差。
与重叠切片相结合,基于 k 空间的重建方法可用于提高多回波 T2SE MRI 的层间分辨率。从 3.2 毫米厚的切片中实现了 1 毫米的分辨率。与标准多回波方法相比,前列腺 MRI 的体内结果显示出更高的锐利度。