College of Life Sciences, Research Center of Analytical Instrumentation, Key Laboratory of Bio-resource and Eco-environment, Ministry of Education, Sichuan University, Chengdu, 610064, Sichuan, China.
Anal Bioanal Chem. 2019 Apr;411(10):2209-2221. doi: 10.1007/s00216-019-01664-x. Epub 2019 Mar 16.
A solid-phase microextraction (SPME) fiber based on a new polymeric ionic liquid was prepared for the extraction of trace aromatic amines in water and their detection by gas chromatography-mass spectrometry (GC-MS). The newly designed polymeric ionic liquid with two functional groups (benzene ring and ether group) was synthesized and fixed on stainless steel wire to effectively extract aromatic amines. Parameters that affect the extraction efficiency of the SPME fiber (extraction temperature, extraction time, alkali concentration, and salt concentration) were optimized to establish a headspace SPME-GC-MS method. The correlation coefficients were 0.996 or greater for concentration of the aromatic amines ranging from 0.01 to 10 μg mL. In addition, the limits of detection for the new fiber are as low as 0.67 ng mL, which is lower than that obtained with polyacrylate. The relative standard deviations of five consecutive extractions for the solution spiked at 1 μg mL by the same fiber were all below 8.3%, and the interfiber relative standard deviations for the solution spiked at the same concentration ranged from 8.9% to 15.2%. Furthermore, long lifetime and good solvent resistance are exhibited by the fiber. Finally, satisfactory relative recovery in the range from 85.3% to 101.9 % was achieved for two environmental water samples.
基于新型聚合离子液体的固相微萃取纤维用于水中痕量芳香胺的萃取及其气相色谱-质谱联用检测。合成了具有两个官能团(苯环和醚基)的新型聚合离子液体,并将其固定在不锈钢丝上,以有效萃取芳香胺。优化了影响 SPME 纤维萃取效率的参数(萃取温度、萃取时间、碱浓度和盐浓度),建立了顶空 SPME-GC-MS 方法。芳香胺浓度在 0.01 至 10μgmL 范围内的相关系数均大于 0.996。此外,新纤维的检出限低至 0.67ngmL,低于聚丙烯酸酯纤维。用同一纤维对 1μgmL 溶液连续 5 次萃取的相对标准偏差均低于 8.3%,同一浓度溶液的纤维间相对标准偏差在 8.9%至 15.2%之间。此外,该纤维还表现出长寿命和好的耐溶剂性。最后,两种环境水样的相对回收率在 85.3%至 101.9%范围内令人满意。