使用调强放射治疗刻度线体模评估瓦里安精确臂和伸缩臂支撑电子射野影像装置的位置准确性。

Evaluation of positional accuracy of the Varian's exact-arm and retractable-arm support electronic portal imaging device using intensity-modulated radiotherapy graticule phantom.

作者信息

Singh Ranjit, Kainth H S, Dev Sachin, Singh Gurjot, Mehta D, Shahi J S, Singh Baljinder, Verma Teerth Raj

机构信息

Department of Radiotherapy, Post Graduate Institute of Medical Education and Research; Department of Physics, Panjab University, Chandigarh, India.

Department of Physics, Panjab University, Chandigarh, India.

出版信息

J Cancer Res Ther. 2019 Jan-Mar;15(1):204-210. doi: 10.4103/jcrt.JCRT_707_17.

Abstract

PURPOSE

The aim of the present study was to compare the positional accuracy of varian's exact-arm (E-arm) and retractable-arm (R-arm) supporting electronic portal imaging device (EPID) systems (amorphous silicon flat-panel detector) using the intensity-modulated radiotherapy (IMRT) graticule phantom.

MATERIALS AND METHODS

The known shifts of 0.5, 1.0, and 1.5 cm were introduced to the given phantom in longitudinal, lateral, and vertical directions, respectively, with respect to treatment couch of medical linear accelerator. The experiment was repeated for different gantry angle and varying source to imager distances (SIDs). The images were acquired for each shift at varying SIDs and beam orientations for both EPID supporting systems. The corresponding shifts obtained from treatment planning system (TPS) were recorded and compared.

RESULTS

The known (expected) and observed (recorded from TPS) shifts obtained for different beam angles (namely, 0°, 90°, 180°, and 270° for anterior, left lateral, posterior, and right-lateral portal images, respectively) in the longitudinal, lateral, and vertical direction at varying SID were compared. The maximum shift in the observed value from the expected one was 3 and 2 mm, respectively, out of the all beam configuration for R-arm and E-arm. These shifts were randomly observed for all imager position and beam orientation.

CONCLUSION

The IMRT graticule phantom is an effective tool to check the mechanical characteristic and consistency of different EPID supporting arms. The effect of EPID sag due to gravity (gantry and treatment couch) was not significant for detection of shift in patient's position. The E-arm support EPID has better mechanical stability and accuracy in detection of patient's position than that of R-arm.

摘要

目的

本研究旨在使用调强放射治疗(IMRT)刻度线体模,比较瓦里安公司的精确臂(E臂)和可伸缩臂(R臂)支撑的电子射野影像装置(EPID)系统(非晶硅平板探测器)的位置精度。

材料与方法

相对于医用直线加速器的治疗床,分别在纵向、横向和垂直方向上对给定体模引入0.5、1.0和1.5厘米的已知位移。针对不同的机架角度和变化的源到成像器距离(SIDs)重复该实验。针对两个EPID支撑系统,在不同的SIDs和射束方向下,对每次位移采集图像。记录并比较从治疗计划系统(TPS)获得的相应位移。

结果

比较了在不同SID下,不同射束角度(即分别对应前野、左侧野、后野和右侧野图像的0°、90°、180°和270°)在纵向、横向和垂直方向上获得的已知(预期)和观察到的(从TPS记录)位移。在所有射束配置中,R臂和E臂观察值与预期值的最大位移分别为3毫米和2毫米。这些位移在所有成像器位置和射束方向上随机观察到。

结论

IMRT刻度线体模是检查不同EPID支撑臂机械特性和一致性的有效工具。重力(机架和治疗床)导致的EPID下垂对检测患者位置的位移影响不显著。与R臂相比,E臂支撑的EPID在检测患者位置方面具有更好的机械稳定性和准确性。