一种用于全景正面光学相干断层扫描血管造影扫描视野量化的新策略。

A novel strategy for quantification of panoramic en face optical coherence tomography angiography scan field.

作者信息

Kadomoto Shin, Uji Akihito, Muraoka Yuki, Akagi Tadamichi, Miyata Manabu, Tsujikawa Akitaka

机构信息

Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan.

出版信息

Graefes Arch Clin Exp Ophthalmol. 2019 Jun;257(6):1199-1206. doi: 10.1007/s00417-019-04310-5. Epub 2019 Apr 10.

Abstract

PURPOSE

To compare the retinal area measured on a panoramic en face optical coherence tomography angiography (OCTA) image with that on an ultra-widefield fluorescein angiography (UWF FA) image.

METHODS

Sixteen eyes (11 with branch retinal vein occlusion, 2 with central retinal vein occlusion, 1 with branch retinal artery occlusion, and 2 with hypertensive retinopathy) were included in this study. A panoramic en face OCTA image was created from five single non-panoramic en face OCTA 12 × 12-mm images. The panoramic OCTA image was superimposed on the corresponding UWF FA image after image registration; the total retinal area was measured using the grid displayed on the UWF FA image. The area on the UWF FA image was measured using stereographic projection software.

RESULTS

The area of retina measured on a single non-panoramic 12 × 12-mm en face OCTA image, a panoramic en face OCTA image, and a UWF FA image was 152.4 ± 3.4 mm, 369.6 ± 26.9 mm, and 813.1 ± 24.4 mm, respectively (P < 0.0001). The panoramic OCTA image was 2.42-fold larger than a single non-panoramic 12 × 12-mm OCTA image and 0.46-fold smaller than a UWF FA image.

CONCLUSIONS

Grid-based measurements using an OCTA image superimposed on the UWF FA image enabled measurement on the panoramic en face OCTA image with minimum influence of magnification errors because of the curved surface of the retina.

摘要

目的

比较全景式正面光学相干断层扫描血管造影(OCTA)图像与超广角荧光素血管造影(UWF FA)图像上测量的视网膜面积。

方法

本研究纳入了16只眼睛(11只患有视网膜分支静脉阻塞,2只患有视网膜中央静脉阻塞,1只患有视网膜分支动脉阻塞,2只患有高血压性视网膜病变)。通过五幅12×12毫米的非全景式正面OCTA单幅图像创建全景式正面OCTA图像。图像配准后,将全景OCTA图像叠加在相应的UWF FA图像上;使用UWF FA图像上显示的网格测量视网膜总面积。使用立体投影软件测量UWF FA图像上的面积。

结果

在单幅非全景式12×12毫米正面OCTA图像、全景式正面OCTA图像和UWF FA图像上测量的视网膜面积分别为152.4±3.4平方毫米、369.6±26.9平方毫米和813.±24.4平方毫米(P<0.0001)。全景OCTA图像比单幅非全景式12×12毫米OCTA图像大2.42倍,比UWF FA图像小0.46倍。

结论

将OCTA图像叠加在UWF FA图像上进行基于网格的测量,能够在全景式正面OCTA图像上进行测量,且由于视网膜曲面的影响,放大误差的影响最小。