Suppr超能文献

用于幕上手术的经颅骨钉直接刺激与经螺旋钻经颅刺激术中运动诱发电位监测的比较。

Comparison of Intraoperative Motor Evoked Potentials Monitoring with Direct Cranial Stimulation by Peg-Screw and Transcranial Stimulation by Corkscrew for Supratentorial Surgery.

作者信息

Kanaya Kohei, Goto Tetsuya, Horiuchi Tetsuyoshi, Hongo Kazuhiro

机构信息

Department of Neurosurgery, Shinshu University School of Medicine, Matsumoto, Japan.

Department of Neurosurgery, Shinshu University School of Medicine, Matsumoto, Japan.

出版信息

World Neurosurg. 2019 Jul;127:e1044-e1050. doi: 10.1016/j.wneu.2019.04.039. Epub 2019 Apr 10.

Abstract

OBJECTIVE

Corkscrew (CS) electrodes are usually used for transcranial electrical stimulation (TES) in the intraoperative monitoring of motor evoked potentials (MEP). Direct cranial stimulation with peg-screw (PS) electrodes can elicit MEP. The present study investigated the difference in the initial threshold between PS and CS electrodes for intraoperative MEP monitoring.

METHODS

We retrospectively analyzed TES-MEP monitoring for supratentorial surgery in 72 patients. Of these 72 patients, 44 were monitored with PS and CS electrodes (PS/CS group) and 28 were monitored with CS and CS electrodes (CS/CS group). TES was used to deliver electrical stimulation by a train of 4-pulse anodal constant current stimulation. The initial threshold in each electrode was checked and analyzed.

RESULTS

In the PS/CS group, the initial threshold with the PS electrode was 38.3 ± 15.1 mA (mean ± standard deviation) on the affected side, and the initial threshold with the CS electrode was 51.4 ± 13.9 mA on the unaffected side. The initial threshold with the PS electrode was significantly lower than that with the CS electrode (P = 0.0001). In the CS/CS group, the initial threshold was 56.2 ± 16.5 mA on the affected side and 62.1 ± 18.6 mA on the unaffected side, with no statistically significant difference (P = 0.23).

CONCLUSION

The initial threshold to elicit MEP was significantly lower with the PS electrode than with the CS electrode. A PS electrode can be used as a feasible stimulation electrode for TES-MEP.

摘要

目的

螺旋(CS)电极通常用于运动诱发电位(MEP)术中监测的经颅电刺激(TES)。用栓钉(PS)电极进行直接颅骨刺激可引出MEP。本研究调查了PS电极和CS电极在术中MEP监测时初始阈值的差异。

方法

我们回顾性分析了72例幕上手术的TES-MEP监测情况。在这72例患者中,44例采用PS电极和CS电极进行监测(PS/CS组),28例采用CS电极和CS电极进行监测(CS/CS组)。通过4脉冲阳极恒流刺激序列使用TES进行电刺激。检查并分析每个电极的初始阈值。

结果

在PS/CS组中,患侧PS电极的初始阈值为38.3±15.1 mA(平均值±标准差),健侧CS电极的初始阈值为51.4±13.9 mA。PS电极的初始阈值显著低于CS电极(P = 0.0001)。在CS/CS组中,患侧初始阈值为56.2±16.5 mA,健侧为62.1±18.6 mA,差异无统计学意义(P = 0.23)。

结论

引出MEP的初始阈值,PS电极显著低于CS电极。PS电极可作为TES-MEP的一种可行刺激电极。