Haemophilia. 2019 May;25(3):e165-e173. doi: 10.1111/hae.13744. Epub 2019 Apr 17.
The joint range of motion (ROM) is an important clinical parameter used to assess the loss of functionality resulting from joint bleedings in people with haemophilia. These episodes require a close follow-up and, to decrease patients' hospital dependence, telemedicine tools are needed. Therefore, this study is aimed to analyse the validity of the Microsoft Kinect V2 sensor with corrected angle measurement to be used in the monitoring of elbow ROM in people with haemophilia.
A convenience sample of 10 healthy controls (CG) and 10 patients with haemophilia with elbow arthropathy (HG) participated in this study. Full ROM of elbow joints was measured in the frontal view with a 10-degree sweep using: (a) a clinical goniometer; (b) the Kinect V2; (c) the Kinect V2 with angle correction; and (d) using a photograph. Bland-Altman graphs (mean and 95% Limits of Agreement [LOA]) and Wilcoxon test were used to determine differences between measurements and groups.
The angle-corrected Kinect V2 measurement removed the skew in the original data, reducing the average errors from 7.9° (LoA = -10.3°; 26.0°; CG) and 9.5° (LoA = -7.9°; 26.9°; HG) to -0.1° (LoA = -8.1°; 7.9°; CG) and -0.7° (LoA = -10.7°; 9.3°; HG).
These error levels allow the use of Kinect V2 in the clinical practice. Kinect V2 with angle correction can complement the classical goniometry allowing an efficient and touchless measurement of ROM.
关节活动度(ROM)是评估血友病患者关节出血导致功能丧失的重要临床参数。这些发作需要密切随访,并且为了减少患者对医院的依赖,需要使用远程医疗工具。因此,本研究旨在分析经角度校正后的 Microsoft Kinect V2 传感器在监测血友病患者肘部 ROM 方面的有效性。
本研究便利选取了 10 名健康对照者(CG)和 10 名患有肘部关节炎的血友病患者(HG)。使用以下方法在额状面测量肘部关节的全 ROM:(a)临床量角器;(b)Kinect V2;(c)带角度校正的 Kinect V2;(d)使用照片。采用 Bland-Altman 图(均值和 95%置信区间 [LOA])和 Wilcoxon 检验来确定测量值和组间的差异。
经角度校正的 Kinect V2 测量去除了原始数据中的偏斜,将平均误差从 7.9°(LOA = -10.3°;26.0°;CG)和 9.5°(LOA = -7.9°;26.9°;HG)降低至-0.1°(LOA = -8.1°;7.9°;CG)和-0.7°(LOA = -10.7°;9.3°;HG)。
这些误差水平允许在临床实践中使用 Kinect V2。经角度校正的 Kinect V2 可以补充传统量角计,实现 ROM 的高效、无接触测量。