Suppr超能文献

Single-Use System Integrity I: Using a Microbial Ingress Test Method to Determine the Maximum Allowable Leakage Limit (MALL).

作者信息

Aliaskarisohi Saeedeh, Hogreve Marc, Langlois Carole, Cutting Jonathan, Barbaroux Magali, Cappia Jean-Marc, Menier Marie-Christine

机构信息

R&D, Sartorius Stedim Biotech GmbH, Goettingen, Germany;

R&D, Sartorius Stedim Biotech GmbH, Goettingen, Germany.

出版信息

PDA J Pharm Sci Technol. 2019 Sep-Oct;73(5):459-469. doi: 10.5731/pdajpst.2018.009688. Epub 2019 Apr 19.

Abstract

An aerosol microbial ingress test was specifically designed and used to create a predictive model in order to determine the maximum allowable leakage limit (MALL) of single-use systems (SUSs). The MALL is defined as the greatest leak size that does not pose any risk to the product. The procedure involved taking test samples of film material from single-use bags. As test samples, an ethylene vinyl acetate multilayer film (300 μm thick) and a polyethylene multilayer film (400 μm thick) were cut into 50 mm patches. Artificial defects of 1-100 µm were laser-drilled in the middle of each film patch. The patch was assembled on a holder and properly sealed. The test units were filled aseptically with culture media and placed inside an aerosol chamber. Various pressures were applied to the test unit to simulate the constraints that single-use systems may be subject to under real-world conditions. After an aerosolization cycle with spores of , a minimum concentration of 10 CFU/cm was reached on the film surface. The test units were incubated for 14 days at 30°C-35°C and visually inspected for bacterial ingress. Thirty samples per defect size were tested. Logistic regression was used to indicate the MALL for a single-use system according to the required risk level. With this method, the probability of the occurrence or absence of ingress for a specific defect size was reported according to the experimental data. In addition to physical parameters, such as the pressure applied and the film material, the effect of the probabilistic nature of the microorganisms in determining the MALL is considered. Although finding an experimental model to predict the MALL for real-life process conditions was the ultimate objective, this paper also presents the microbial ingress test data obtained so far for two extreme conditions. Potential constraints, such as vibration, shock, acceleration, liquid movement, and pressure differentials, observed during normal usage were simulated using two extreme differential pressures, 0 mbar and 300 mbar. The estimated MALL for typical use-case conditions are 10-20 µm for storage applications and 2-10 µm for shipping conditions. The microbial integrity test method used in this article was able to detect bacterial ingress down to 3 µm defect size. As use of single-use systems (SUSs) is increasingly expanding into all process steps of commercial manufacturing, integrity failure can significantly impact drug safety, availability, and costs. Consequently, growing industry scrutiny on single-use system integrity (SUSI) is raising the need to develop good science behind reliable determinations of liquid leakage and microbial ingress as well as the appropriate physical integrity testing technologies. In the current study, microbial ingress testing by the aerosol method is used to determine the maximum allowable leakage limit (MALL) for SUSs. To define the MALL, it is generally assumed that a system or product will not show any microbial ingress or leakage at a certain defect size. Statistical analysis of the experimental data in this study indicated the MALL with probability at a certain defect size for each system. As a result, the method studied provides a more accurate way of predicting ingress and increasing safety down the line for drug manufacturers and patients alike.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验