Suppr超能文献

在不到 3 分钟内以 1 毫米各向同性分辨率实现前列腺的加速 3D T w 成像。

Accelerated 3D T w-imaging of the prostate with 1-millimeter isotropic resolution in less than 3 minutes.

机构信息

School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom.

MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom.

出版信息

Magn Reson Med. 2019 Aug;82(2):721-731. doi: 10.1002/mrm.27764. Epub 2019 Apr 21.

Abstract

PURPOSE

To achieve 3D T w imaging of the prostate with 1-mm isotropic resolution in less than 3 min.

METHODS

We devised and implemented a 3D T -prepared multishot balanced steady state free precession (T prep-bSSFP) acquisition sequence with a variable density undersampled trajectory combined with a total variation regularized iterative SENSE (TV-SENSE) reconstruction. Prospectively undersampled images of the prostate (acceleration factor R = 3) were acquired in 11 healthy subjects in an institutional review board-approved study. Image quality metrics (subjective signal-to-noise ratio, contrast, sharpness, and overall prostate image quality) were evaluated by 2 radiologists. Scores of the proposed accelerated sequence were compared using the Wilcoxon signed-rank and Kruskal-Wallis non-parametric tests to prostate images acquired using a fully sampled 3D T prep-bSSFP acquisition, and with clinical standard 2D and 3D turbo spin echo (TSE) T w acquisitions. A P-value < 0.05 was considered significant.

RESULTS

The 3× accelerated 3D T prep-bSSFP images required a scan time (min:s) of 2:45, while the fully sampled 3D T prep-bSSFP and clinical standard 3D TSE images were acquired in 8:23 and 7:29, respectively. Image quality scores (contrast, sharpness, and overall prostate image quality) of the accelerated 3D T prep-bSSFP, fully sampled T prep-bSSFP, and clinical standard 3D TSE acquisitions along all 3 spatial dimensions were not significantly different (P > 0.05).

CONCLUSION

3D T w images of the prostate with 1-mm isotropic resolution can be acquired in less than 3 min, with image quality that is comparable to a clinical standard 3D TSE sequence but only takes a third of the acquisition time.

摘要

目的

实现前列腺 1 毫米各向同性分辨率的 3D T w 成像,扫描时间少于 3 分钟。

方法

我们设计并实现了一种 3D T 准备多shot 平衡稳态自由进动(T prep-bSSFP)采集序列,该序列采用可变密度欠采样轨迹,结合全变分正则化迭代 SENSE(TV-SENSE)重建。在一项机构审查委员会批准的研究中,对 11 名健康受试者的前列腺进行了前瞻性欠采样图像采集(加速因子 R = 3)。由 2 名放射科医生评估图像质量指标(主观信噪比、对比度、锐利度和整体前列腺图像质量)。使用 Wilcoxon 符号秩检验和 Kruskal-Wallis 非参数检验对加速序列评分与完全采样 3D T prep-bSSFP 采集、临床标准 2D 和 3D 涡轮自旋回波(TSE)T w 采集的前列腺图像进行比较。P 值<0.05 被认为有统计学意义。

结果

3 倍加速的 3D T prep-bSSFP 图像需要 2:45 分钟,而完全采样的 3D T prep-bSSFP 和临床标准 3D TSE 图像分别需要 8:23 分钟和 7:29 分钟。加速 3D T prep-bSSFP、完全采样 T prep-bSSFP 和临床标准 3D TSE 采集的图像质量评分(对比度、锐利度和整体前列腺图像质量)在所有 3 个空间维度上均无显著差异(P > 0.05)。

结论

可以在不到 3 分钟的时间内获得 1 毫米各向同性分辨率的前列腺 3D T w 图像,其图像质量与临床标准的 3D TSE 序列相当,但采集时间仅为其三分之一。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d196/6563534/0a08014ab575/MRM-82-721-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验