Stacchi Claudio, Matteo De Biasi, Torelli Lucio, Robiony Massimo, Di Lenarda Roberto, Angerame Daniele
1 TRIESTE UNIVERSITY BIOMEDICINE University of Trieste.
2 Universita degli Studi di Trieste Research Fellow University Clinical Department of Medical, Surgical and Health Sciences University of Trieste.
J Oral Implantol. 2019 Apr 22. doi: 10.1563/aaid-joi-D-18-00157.
The primary objective of the present in vitro study was to evaluate the influence of implant site preparation technique (drills vs. ultrasonic instrumentation) on the primary stability of short dental implants with two different designs inserted in simulated low quality cancellous bone. Eighty implant sites were prepared in custom-made solid rigid polyurethane blocks with two different low cancellous bone densities (5 or 15 pounds per cubic foot, PCF), equally distributed between piezoelectric (Surgysonic Moto, Esacrom, Italy) and conventional drilling techniques. Two short implant systems (Prama and Syra, Sweden & Martina) were tested by inserting 40 fixtures of each system (both 6.0 mm length and 5.0 mm diameter), divided in the four subgroups (drills/5 PCF density; drills/15 PCF density; piezo/5 PCF density; piezo/15 PCF density). Insertion torque (Ncm), implant stability quotient values, removal torque (Ncm) and surgical time were recorded. Data were analyzed by three-way ANOVA and Scheffé's test (α=0.05). With slight variations among the considered dependent variables, overall high primary implant stability was observed across all subgroups. Piezoelectric instrumentation allowed for comparable or slightly superior primary stability in comparison with the drilling procedures in both implant systems. Prama implants group showed the highest mean reverse torque, Syra implants the highest implant stability quotient values. Piezoelectric implant site preparation took prolonged operative time compared to conventional preparation with drills; among the drilling procedures, Syra system requires fewer surgical steps and shorter operative time.
本体外研究的主要目的是评估种植位点预备技术(钻针与超声器械)对两种不同设计的短牙种植体在模拟低质量松质骨中植入时初期稳定性的影响。在定制的固体刚性聚氨酯块中制备了80个种植位点,这些聚氨酯块具有两种不同的低松质骨密度(每立方英尺5磅或15磅,PCF),在压电(意大利Esacrom公司的Surgysonic Moto)和传统钻针技术之间平均分配。测试了两种短种植体系统(瑞典&玛缇娜公司的Prama和Syra),每种系统植入40枚种植体(长度均为6.0 mm,直径均为5.0 mm),分为四个亚组(钻针/5 PCF密度;钻针/15 PCF密度;压电/5 PCF密度;压电/15 PCF密度)。记录植入扭矩(Ncm)、种植体稳定性商值、取出扭矩(Ncm)和手术时间。数据采用三因素方差分析和谢费检验(α = 0.05)进行分析。在所考虑的因变量之间存在轻微差异,所有亚组均观察到总体较高的种植体初期稳定性。与两种种植体系统中的钻针程序相比,压电器械的初期稳定性相当或略优。Prama种植体组的平均反向扭矩最高,Syra种植体的种植体稳定性商值最高。与传统钻针预备相比,压电种植位点预备的手术时间延长;在钻针程序中,Syra系统需要的手术步骤更少,手术时间更短。