Koper Maarten, Verdijk Rob, Bos Koen
Department of Orthopedics, Erasmus Medical Center, Rotterdam, the Netherlands.
Department of Pathology, Erasmus Medical Center, Rotterdam, the Netherlands.
Arthroplast Today. 2019 Jan 22;5(1):38-42. doi: 10.1016/j.artd.2018.12.001. eCollection 2019 Mar.
With the increased use of dual-mobility cups (DMCs) in total-and revision hip arthroplasties, surgeons can expect an increase of known and new complications. During routine follow-up, we observed an asymptomatic patient with an intraprosthetic-dislocation (IPD) and elevated levels of serum metal ions (1.8 ppb of cobalt and 28.0 ppb of chromium). Revision surgery was inevitable. Perioperative metallosis and severe wear of the metal shell and metal femoral head supported the IPD. Literature showed that the modularity of the DMC can result in increased serum metal ions, create excessive wear, and possibly affect implant survival. Our case and review of the literature may form an argument not to consider DMC for primary cases. Furthermore, we advise yearly clinical and radiological follow-up and, on indication, metal ion testing for DMCs.
随着双动杯(DMC)在初次和翻修髋关节置换术中的使用增加,外科医生可能会面临更多已知和新出现的并发症。在常规随访期间,我们观察到一名无症状患者发生了假体内部脱位(IPD),血清金属离子水平升高(钴为1.8 ppb,铬为28.0 ppb)。翻修手术不可避免。围手术期金属沉着症以及金属杯和金属股骨头的严重磨损支持了IPD的诊断。文献表明,DMC的模块化可能导致血清金属离子增加、产生过度磨损,并可能影响植入物的使用寿命。我们的病例及文献回顾可能支持不将DMC用于初次手术病例的观点。此外,我们建议对DMC进行每年一次的临床和影像学随访,并根据指征进行金属离子检测。