During 1984-85, users of a solid-phase radioimmunoassay kit for LH and FSH had problems of both variability and bias in the assay of some EQAS specimens despite adequate internal QC. The cause has been identified as contamination of these specimens with Pseudomonas fluorescens.
在1984年至1985年期间,用于促黄体生成素(LH)和促卵泡生成素(FSH)的固相放射免疫分析试剂盒的用户,尽管进行了充分的内部质量控制,但在一些外部质量评估计划(EQAS)标本的检测中仍存在变异性和偏差问题。原因已确定为这些标本被荧光假单胞菌污染。