Suppr超能文献

不同接触应力下 CFR PEEK 关节对 CoCr 的磨损行为:CFR PEEK 的低磨损被 CoCr 对磨面的磨损所抵消。

Wear behaviour of CFR PEEK articulated against CoCr under varying contact stresses: Low wear of CFR PEEK negated by wear of the CoCr counterface.

机构信息

School of Engineering, Newcastle University, Newcastle Upon Tyne, UK.

School of Engineering, Newcastle University, Newcastle Upon Tyne, UK.

出版信息

J Mech Behav Biomed Mater. 2019 Sep;97:117-125. doi: 10.1016/j.jmbbm.2019.05.022. Epub 2019 May 15.

Abstract

Total hip replacement with metal-on-polymer hip prostheses is the most common treatment for late-stage osteoarthritis. However, the wear debris generated from the polymer acetabular liner remains a problem. Alternative materials with claimed superior wear properties have been proposed to overcome this problem. In this study, the wear behaviour of carbon fibre reinforced polyether ether ketone (CFR PEEK) was investigated under different contact stresses that are observed in the natural hip joint. A 50-station pin-on-disc machine (SuperCTPOD) was used to investigate the wear behaviour of 50 CFR PEEK pins articulated against cobalt chromium (CoCr) discs under five different contact stresses, namely 1.11, 1.38, 1.61, 2.00 and 5.30 MPa. The results showed that the wear rates of the pins did not differ significantly between groups under different contact stresses. In addition, CFR PEEK produced lower wear rates than ultrahigh molecular weight polyethylene and cross-linked polyethylene. However, the weight of the CoCr discs was found to decrease significantly at the end of the wear test, which was indicative of metallic wear. The findings of this study indicated that, despite having relatively low wear rates, CFR PEEK is not a good alternative to be utilised against orthopaedic metals.

摘要

全髋关节置换术采用金属对聚合物髋关节假体是治疗晚期骨关节炎最常用的方法。然而,聚合物髋臼衬垫产生的磨损碎片仍然是一个问题。已经提出了具有声称的优越磨损性能的替代材料来克服这个问题。在这项研究中,研究了在自然髋关节中观察到的不同接触应力下碳纤维增强聚醚醚酮(CFR PEEK)的磨损行为。使用 50 工位销盘机(SuperCTPOD)研究了在五个不同接触应力下,即 1.11、1.38、1.61、2.00 和 5.30 MPa 下,50 个 CFR PEEK 销与钴铬(CoCr)盘之间的磨损行为。结果表明,在不同接触应力下,销的磨损率在各组之间没有显著差异。此外,CFR PEEK 的磨损率低于超高分子量聚乙烯和交联聚乙烯。然而,在磨损试验结束时,发现 CoCr 盘的重量明显下降,表明存在金属磨损。这项研究的结果表明,尽管 CFR PEEK 的磨损率相对较低,但它并不是对抗骨科金属的好替代品。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验