Scalabre Aurélien, Maniouloux Floriane, Vermersch Sophie, Patoir Arnaud, Haddad Elie, Lopez Manuel, Varlet François, Tiffet Olivier
Chest Wall Deformities Unit, University Hospital of Saint-Etienne, CHU de Saint-Etienne Hôpital Nord, Saint Etienne Cedex 2, France.
Interact Cardiovasc Thorac Surg. 2019 Oct 1;29(4):503-509. doi: 10.1093/icvts/ivz145.
The OrtenBodyOne scanner is a radiation-free, 3-dimensional imaging system recently developed for evaluation of the severity of pectus excavatum (PE). The goal of this study was to evaluate the utility of this new imaging system compared with that of computed tomography (CT) for the evaluation of the severity of PE.
Patients treated for PE from April 2015 to January 2017 with available CT and OrtenBodyOne data were included. Correlations between indexes calculated from CT and from OrtenBodyOne were determined by applying the non-parametric Spearman correlation procedure with a Bonferroni correction to adjust for multiple comparisons.
Forty men (90.9%) and 4 women (9.1%), 20 with symmetrical (45.5%) and 24 with asymmetrical PE (54.5%), were included. The median age was 16.1 years (range 4.3-63.5 years). The following measures and indexes acquired using OrtenBodyOne and CT were significantly correlated: pectus depth (r = 0.84; P = 0.002), anthropometric index (r = 0.81; P = 0.002) and asymmetry index (r = 0.67; P = 0.002). The correlation between the CT Haller index and the external Haller index was only significant for symmetrical PE (r = 0.57; P = 0.008).
The OrtenBodyOne imaging system can be used to evaluate the severity of symmetrical PE using the external Haller index. Asymmetry and anthropometric indexes are more reliable for the evaluation of asymmetrical PE. Measures can be repeated throughout treatment while avoiding unnecessary irradiation.
OrtenBodyOne扫描仪是一种最近开发的无辐射三维成像系统,用于评估漏斗胸(PE)的严重程度。本研究的目的是评估这种新成像系统与计算机断层扫描(CT)相比在评估PE严重程度方面的效用。
纳入2015年4月至2017年1月期间接受PE治疗且有可用CT和OrtenBodyOne数据的患者。通过应用非参数Spearman相关程序并进行Bonferroni校正以调整多重比较,确定CT和OrtenBodyOne计算的指标之间的相关性。
纳入40名男性(90.9%)和4名女性(9.1%),其中20名患有对称性PE(45.5%),24名患有非对称性PE(54.5%)。中位年龄为16.1岁(范围4.3 - 63.5岁)。使用OrtenBodyOne和CT获得的以下测量值和指标显著相关:漏斗胸深度(r = 0.84;P = 0.002)、人体测量指数(r = 0.81;P = 0.002)和不对称指数(r = 0.67;P = 0.002)。CT哈勒指数与外部哈勒指数之间的相关性仅在对称性PE中显著(r = 0.57;P = 0.008)。
OrtenBodyOne成像系统可用于使用外部哈勒指数评估对称性PE的严重程度。不对称和人体测量指数在评估非对称性PE时更可靠。在整个治疗过程中可以重复测量,同时避免不必要的辐射。