Lionberger David, Conlon Christopher, Wattenbarger Laura, Walker Timothy J
Southwest Orthopedic Group, The Methodist Hospital at Houston, Houston, TX, USA.
Texas A&M College of Medicine, The Methodist Hospital at Houston, Houston, TX, USA.
Arthroplast Today. 2019 Mar 22;5(2):187-192. doi: 10.1016/j.artd.2019.02.002. eCollection 2019 Jun.
Aseptic loosening is one of the most elusive problems in total knee arthroplasty. We compared the failure rates of posterior cruciate-substituting total knee arthroplasty utilizing implants with hardened surface coating to a previous cohort of patients who underwent the procedure with traditional cruciate-retaining noncoated cobalt-chrome implants.
A retrospective study was conducted of 1099 total knee arthroplasties performed from 2009 to 2017. Two hundred forty-nine total knee arthroplasties performed from January 2015 to March 2017 under a single design were reviewed retrospectively and compared to the author's previous 850 total knee arthroplasties performed from January 2009 to December 2014 under a different design.
This series demonstrated an alarming debonding of cement in the tibial implant. The resultant failure rate of 6% ( < .001) is higher than observed in 850 total knee arthroplasties in the previous 5 years and higher than those reported in the literature giving cause for concern regarding this implant.
Due to the observed excessive failure rate, the authors recommend exercising high levels of caution using this implant with hardened surface treatment until further testing can be ascertained as to the root cause of failure.
无菌性松动是全膝关节置换术中最难以捉摸的问题之一。我们将使用表面硬化涂层植入物的后交叉韧带替代型全膝关节置换术的失败率与先前一组接受传统保留交叉韧带的非涂层钴铬合金植入物手术的患者进行了比较。
对2009年至2017年进行的1099例全膝关节置换术进行了回顾性研究。回顾性分析了2015年1月至2017年3月在单一设计下进行的249例全膝关节置换术,并与作者先前在2009年1月至2014年12月在不同设计下进行的850例全膝关节置换术进行了比较。
该系列显示胫骨植入物中骨水泥出现了令人担忧的脱粘现象。由此产生的6%的失败率(<0.001)高于前5年850例全膝关节置换术中观察到的失败率,也高于文献报道的失败率,这引发了对这种植入物的担忧。
由于观察到的过高失败率,作者建议在确定失败的根本原因进行进一步测试之前,在使用这种表面硬化处理的植入物时要格外谨慎。