采用便携式测力计仰卧位锚固系统测量膝关节伸肌力量的可靠性和有效性。

Reliability and validity of knee extensor strength measurements using a portable dynamometer anchoring system in a supine position.

机构信息

Department of Rehabilitation Medicine, Seoul National University Hospital, Seoul, Republic of Korea.

Department of Rehabilitation Medicine, Seoul National University Hospital, Seoul National University College of Medicine, 101 Daehak-Ro, Jongno-Gu, Seoul, 03080, Republic of Korea.

出版信息

BMC Musculoskelet Disord. 2019 Jul 8;20(1):320. doi: 10.1186/s12891-019-2703-0.

Abstract

BACKGROUND

Muscle strength measurements using hand-held dynamometry (HHD) can be affected by the inadequate strength of the tester and lack of stabilization of the participants and the device. A portable HHD anchoring system was designed that enabled the measurement of isometric knee extensor muscle strength in a supine position. This can be used with individuals who are unable to assume the sitting position required for the measurement of knee extensor strength in conventional isokinetic dynamometry (IKD). The aim of this study was to evaluate the reliability and validity of knee extensor strength measurements using this device.

METHODS

The maximal knee extensor isometric strength of the dominant leg in healthy adults aged 20 to 40 years was tested. Three trials of three contractions were assessed by two raters using the portable dynamometer anchoring system whilst the participant was in the supine position. After the three measurement trials, peak knee extensor torque was evaluated using IKD. The intraclass correlation coefficient (ICC) and 95% limits of agreement (LOA) for intra- and inter-rater reliability were obtained.

RESULTS

Thirty-nine participants (19 male and 20 female, aged 30.08 ± 4.16 y), completed the three measurement trials. The ICC for intra-rater reliability was 0.98 for the maximum measurements of knee extensor strength (95% confidence interval [CI]: 0.96-0.98) and 0.98 (95% CI: 0.96-0.99) for inter-rater reliability. The mean difference (%) between the maximum knee extensor strength measurements of each trial was 1.02% (LOA range: - 11.13 to 13.16%) for intra-rater and - 1.44% (LOA range: - 13.98 to 11.08%) for inter-rater measurements. The Pearson correlation coefficient of the maximum voluntary peak torque measurements with the portable dynamometer anchoring system and IKD was 0.927.

CONCLUSIONS

The portable dynamometer anchoring system is a reliable and valid tool for measuring isometric knee extensor strength in a supine position. Future clinical feasibility studies are needed to determine if this equipment can be applied to people with severe illness or disabilities.

TRIAL REGISTRATION

KCT0003041 .

摘要

背景

使用手持测力计(HHD)进行肌肉力量测量可能会受到测试者力量不足以及参与者和设备不稳定的影响。设计了一种便携式 HHD 锚固系统,该系统可用于测量仰卧位时的等长膝关节伸肌肌力。这可用于无法采取传统等速测力计(IKD)测量膝关节伸肌力量所需的坐姿的个体。本研究的目的是评估该设备测量膝关节伸肌力量的可靠性和有效性。

方法

测试了 20 至 40 岁健康成年人的优势腿最大膝关节等长伸肌力量。两名评估者使用便携式测力计锚固系统,让参与者处于仰卧位,评估三次三次收缩的三次试验。在三次测量试验后,使用 IKD 评估峰值膝关节伸肌扭矩。获得了内部和内部评估者之间可靠性的组内相关系数(ICC)和 95%置信区间(CI)的协议范围(LOA)。

结果

39 名参与者(19 名男性和 20 名女性,年龄 30.08 ± 4.16 岁)完成了三次测量试验。内评估者可靠性的 ICC 为 0.98,用于膝关节伸肌强度的最大测量(95%CI:0.96-0.98),而内评估者可靠性的 ICC 为 0.98(95%CI:0.96-0.99)。每个试验的最大膝关节伸肌强度测量之间的平均差异(%)为 1.02%(LOA 范围:-11.13 至 13.16%)用于内部评估者和-1.44%(LOA 范围:-13.98 至 11.08%)用于内部评估者测量。与便携式测力计锚固系统和 IKD 的最大自愿峰值扭矩测量的 Pearson 相关系数为 0.927。

结论

便携式测力计锚固系统是一种可靠且有效的工具,可用于测量仰卧位时的等长膝关节伸肌强度。需要进行未来的临床可行性研究,以确定该设备是否可用于患有严重疾病或残疾的人群。

试验注册

KCT0003041。