The Amplatzer vascular plugs (AVPs), including AVP 4, have excellent occlusion; however, insufficient embolization or recanalization may occasionally occur. This study aimed to evaluate the feasibility and effectiveness of internal coil packing method to insert microcoils into the AVP 4 using an experimental vascular model. The insertion of a 1.7 F microcatheter through the nitinol mesh of the AVP 4 was not possible. Only 0.010-inch electrolytic detachable microcoils could be inserted through catheter tips in contact with nitinol mesh. The internal coil packing of the AVP 4 may be accomplished by inserting 0.010-inch microcoils into the AVP 4 in cases of continued perfusion or recanalization with required short-segment embolization.
Amplatzer 血管塞(AVP),包括 AVP4,具有极好的闭塞效果;然而,偶尔可能会出现栓塞不足或再通。本研究旨在评估使用实验性血管模型将微线圈插入 AVP4 中的内部线圈包装方法的可行性和有效性。通过 AVP4 的镍钛诺网插入 1.7F 微导管是不可能的。只有与镍钛诺网接触的导管尖端才能插入 0.010 英寸电解可分离微线圈。如果需要进行短节段栓塞的持续灌注或再通,则可以通过将 0.010 英寸微线圈插入 AVP4 来完成 AVP4 的内部线圈包装。