Holburn A M, Prior D M, Whitton C M
Blood Group Reference Laboratory, Oxford, UK.
Clin Lab Haematol. 1988;10(1):73-85. doi: 10.1111/j.1365-2257.1988.tb01155.x.
In seven exercises of blood grouping the overall rates of major error were 0.19% and 0.25% in ABO and D grouping respectively. In ABO grouping this represents an increase in error rate over that observed in 1982-1983 but the increase was due to an unusually high error rate with one particular group A2B cell. An improvement in performance was observed in simple D grouping and was largely due to a lower incidence of false positive grouping of D-negative cells in the antiglobulin test. An improvement in performance observed in D grouping IgG-coated D-negative cells appeared to be due to a better understanding of the problem rather than to any change in serological practice. Error rates in antibody screening were somewhat lower than in 1982-1983 but this may or may not represent an improvement in performance as the test materials were not the same in the two periods. The direct antiglobulin test with IgG-coated cells was reliably performed with polyspecific and with anti-IgG reagents but an excess of false positive results was obtained with anti-C3d. Error rates in antibody identification varied from 0.6% for anti-D to 74% for anti-c + E.
在七项血型检测中,ABO血型和D血型检测的主要错误总体发生率分别为0.19%和0.25%。在ABO血型检测中,这一错误率相较于1982 - 1983年有所上升,但上升是由于一个特定的A2B细胞出现了异常高的错误率。在简单D血型检测中观察到性能有所改善,这主要归因于抗球蛋白试验中D阴性细胞假阳性分组的发生率较低。在D血型检测中,针对IgG包被的D阴性细胞观察到的性能改善似乎是由于对该问题有了更好的理解,而非血清学操作有任何改变。抗体筛查的错误率略低于1982 - 1983年,但由于两个时期的检测材料不同,这可能代表也可能不代表性能的改善。用IgG包被细胞进行的直接抗球蛋白试验使用多特异性试剂和抗IgG试剂时结果可靠,但使用抗C3d时获得了过多的假阳性结果。抗体鉴定的错误率从抗D的0.6%到抗c + E的74%不等。