Suppr超能文献

采用波长色散X射线荧光光谱法对新型有证标准物质进行瓶间均匀性测试。

Between-bottle homogeneity test of new certified reference materials employing wavelength dispersive X-ray fluorescence spectrometry.

作者信息

Idris Abubakr M

机构信息

1Department of Chemistry, College of Science, King Khalid University, Abha, 61321 Saudi Arabia.

2Research Center for Advanced Materials Science (RCAMS), King Khalid University, P.O. Box 9004, Abha, 61413 Saudi Arabia.

出版信息

BMC Chem. 2019 Feb 12;13(1):23. doi: 10.1186/s13065-019-0528-4. eCollection 2019 Dec.

Abstract

BACKGROUND

A pilot study has being carried out at our laboratories, with international collaborators, to develop seven certified reference materials (CRMs), which have matrices of mainly soil and biological tissues. The CRMs will be certified for macro-, micro- and trace elements for environmental, toxicological, agronomic and nutritional purposes. Homogeneity of element concentrations is a critical step in the production process of CRMs. This work employs wavelength dispersive X-ray fluorescence spectrometry (WD-XRF) to test between-bottle homogeneity of the CRMs.

RESULTS

The relative standard deviation (RSD), relative average deviation, Skewness and Kurtosis of element contents in seven bottles out of 80 bottles of each CRM were considered to assess homogeneity in terms of variability and distribution. More than 50% of the number of quantified elements recorded RSD between 2 and 5%. Hence, based on an in-house classification, the intended CRMs recorded excellent to good between-bottle homogeneity. Nevertheless, the contents of some elements (Ni, Rb, Zn and Br) experienced high RSD values (> 10%). The Skewness and Kurtosis values of most elements are around one indicating symmetric distribution and thus have an absence of tailing relative to the normal distribution.

CONCLUSIONS

WD-XRF provides fit for purpose data for assessment of initial between-bottle homogeneity in terms of rapidity, ease of use, multi-element quantification and sample non-destruction.

摘要

背景

我们的实验室与国际合作者开展了一项试点研究,以开发七种有证标准物质(CRM),其基质主要为土壤和生物组织。这些CRM将针对环境、毒理学、农艺学和营养目的的常量、微量和痕量元素进行定值。元素浓度的均匀性是CRM生产过程中的关键步骤。本研究采用波长色散X射线荧光光谱法(WD-XRF)来检测CRM瓶间均匀性。

结果

在每种CRM的80瓶样品中,选取7瓶的元素含量的相对标准偏差(RSD)、相对平均偏差、偏度和峰度,从变异性和分布角度评估均匀性。超过50%的定量元素的RSD在2%至5%之间。因此,根据内部分类,预期的CRM记录的瓶间均匀性为优至良。然而,某些元素(镍、铷、锌和溴)的含量的RSD值较高(>10%)。大多数元素的偏度和峰度值约为1,表明分布对称,因此相对于正态分布没有拖尾现象。

结论

WD-XRF在快速性、易用性、多元素定量和样品无损方面提供了适用于评估初始瓶间均匀性的数据。