Performance characteristics of a whole-body positron tomograph.

作者信息

Spinks T J, Guzzardi R, Bellina C R

机构信息

MRC Cyclotron Unit, Hammersmith Hospital, London, UK.

出版信息

J Nucl Med. 1988 Nov;29(11):1833-41.

PMID:3141597
Abstract

This paper describes an investigation of some of the important physical characteristics of a whole-body positron tomograph consisting of two rings of bismuth germanate detectors of dimensions 5.6 mm X 30 mm X 30 mm (512/ring). The resolution applicable to in vivo imaging is six mm or more, depending on radionuclide and reconstruction filter and is very uniform over the field of view normally used. Axial resolution can be varied by moving side collimation (maximum approximately 16 mm FWHM with interplane septa). A simple scheme has been devised to correct for loss of true coincidence events with varying count rate based on the total random and multiple coincidence rates. It is concluded that correction for scattered radiation should be implemented for reliable quantitation.

摘要

本文描述了对一台全身正电子断层扫描仪一些重要物理特性的研究,该扫描仪由两环锗酸铋探测器组成,探测器尺寸为5.6毫米×30毫米×30毫米(每环512个)。适用于体内成像的分辨率为6毫米或更高,这取决于放射性核素和重建滤波器,并且在通常使用的视野范围内非常均匀。轴向分辨率可通过移动侧向准直来改变(使用平面间隔时,半高宽最大约为16毫米)。已经设计了一种简单的方案,根据总随机符合率和多重符合率来校正随计数率变化的真符合事件损失。得出的结论是,为了进行可靠的定量分析,应实施散射辐射校正。