Analyst. 2019 Sep 9;144(18):5449-5454. doi: 10.1039/c9an01087d.
A simple-to-use device to measure drugs in saliva, blood plasma, and whole blood for point-of-care analysis and treatment of overdose patients has been investigated. A rudimentary flow strip has been developed to separate opioids from these biofluids for analysis by surface-enhanced Raman spectroscopy (SERS). The strips are based on lateral flow assays, in which the antibodies have been substituted by SERS-active pads for detection. Samples of codeine and fentanyl, artificially added to these biofluids, were measured using the strips by a field-usable Raman spectrometer. We report measurement of these drugs in these biofluids from 0.5 to 5 μg mL-1 in 5 minutes. Calculated limits of detection for the spectra suggest that these drugs could be measured at 5 to 20 ng mL-1 with improvements in the strips' separation capability.