Suppr超能文献

[利用黄金射束数据对放射治疗计划系统进行高效调试]

[Efficient Commissioning of the Radiotherapy Treatment Planning System with the Golden Beam Data].

作者信息

Hiroki Tomoyuki, Fujita Yukio, Maehira Shota, Soda Kenji, Kikuchi Tomoko, Kunieda Etsuo

机构信息

Department of Radiology, Tokai University Hospital.

Department of Radiation Oncology, Tokai University School of Medicine (Current address: Department of Radiological Sciences, Komazawa University).

出版信息

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2019;75(8):725-735. doi: 10.6009/jjrt.2019_JSRT_75.8.725.

Abstract

Commissioning of a linear accelerator (Linac) and treatment planning systems (RTPs) for clinical use is complex and time-consuming, typically 3-4 months in total. However, based on clinical needs and economics, hospitals desire early clinical starts for patients, and various studies have been conducted for shortening the preparation period. One of the methods to shorten the period is using golden beam data (GBD). The purpose of this study was to shorten the commissioning period without reducing accuracy and to simplify commissioning works while improving safety. We conducted commissioning of the RTPs before installing the Linac using GBD, and carried out verification immediately after the acceptance test. We used TrueBeam STx (Varian Medical Systems) and Eclipse (ver. 13.7, Varian Medical Systems) for RTPs and anisotropic analysis algorithm (AAA) and AcurosXB (AXB) for calculation algorithms. The difference between GBD and the measured beam data was 0.0 ± 0.2% [percentage depth dose (PDDs) ] and -0.1 ± 0.2% (Profiles) with X-ray, and -1.2 ± 1.3% (PDDs) with electrons. The difference between the calculated dose and the measured dose was 0.1 ± 0.3% (AAA) and 0.0 ± 0.3% (AXB) under homogeneous conditions, and 0.7 ± 1.4% (AAA) and 0.6 ± 1.1% (AXB) under heterogeneous conditions. We took 43 days from the end of the acceptance test to the start of clinical use. We found that the preparation period for clinical use can be shortened without reducing the accuracy, by thinning out the number of measurement items using GBD.

摘要

用于临床的直线加速器(Linac)和放射治疗计划系统(RTPs)的调试复杂且耗时,总共通常需要3 - 4个月。然而,基于临床需求和经济性,医院希望能尽早为患者开展临床治疗,并且已经进行了各种研究以缩短准备期。缩短该时期的方法之一是使用黄金束数据(GBD)。本研究的目的是在不降低准确性的情况下缩短调试期,并在提高安全性的同时简化调试工作。我们在安装直线加速器之前使用GBD对RTPs进行调试,并在验收测试后立即进行验证。我们使用TrueBeam STx(瓦里安医疗系统公司)和Eclipse(版本13.7,瓦里安医疗系统公司)作为RTPs,使用各向异性分析算法(AAA)和AcurosXB(AXB)作为计算算法。GBD与实测束数据之间的差异,对于X射线,百分深度剂量(PDDs)为0.0±0.2%,剂量分布曲线为 - 0.1±0.2%;对于电子,PDDs为 - 1.2±1.3%。在均匀条件下,计算剂量与实测剂量之间的差异,AAA为0.1±0.3%,AXB为0.0±0.3%;在非均匀条件下,AAA为0.7±1.4%,AXB为0.6±1.1%。从验收测试结束到临床使用开始我们花了43天。我们发现,通过使用GBD精简测量项目数量,可以在不降低准确性的情况下缩短临床使用的准备期。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验