细胞因子在体外血液净化中的清除:一项体外研究。
Cytokine Removal in Extracorporeal Blood Purification: An in vitro Study.
机构信息
Department for Biomedical Research, Danube University Krems, Krems, Austria,
Department of Pharmaceutical Technology and Biopharmaceutics, University of Vienna, Vienna, Austria,
出版信息
Blood Purif. 2020;49(1-2):33-43. doi: 10.1159/000502680. Epub 2019 Sep 11.
BACKGROUND/AIMS: Cytokines are among the main target substances that have to be removed effectively in order to improve the patient's health status in the treatment of sepsis, septic shock, and liver diseases. Although there are various medical devices commercially available, the success of their clinical use is limited. The aim of this in vitro study was to compare 3 different medical devices with respect to their clearance for the cytokines interleukin-6 (IL-6), IL-8, IL-1β, and tumor necrosis factor alpha. The medical devices that were tested are the whole blood adsorbent CytoSorb, the high cutoff filter EMiC2, and the hemofilter HemofeelCH 1.8.
METHODS
The study was carried out on the multiFiltrate machine with 1 L human plasma for 8 h. Samples for cytokine quantification were taken at defined time points from the plasma pool. Each experiment was conducted in triplicates, and clearance was calculated for all tested cytokines.
RESULTS
All 3 medical devices showed good cytokine removal. The highest clearance for all cytokines was achieved by hemoperfusion with Cytosorb. IL-8 and IL-6 clearance were higher with Hemofeel (continuous venovenous hemodiafiltration) than with EMiC2 (continuous venovenous hemodialysis) because the polymethyl methacrylate (PMMA)-based membrane Hemofeel is able to remove these 2 cytokines by adsorption. Protein and albumin loss was highest withCytosorb and lowest with EMiC2.
CONCLUSION
The mechanisms of cytokine removal by blood purification include convection, diffusion, and adsorption. PMMA-based filters are able to combine all 3 mechanisms for certain cytokines. Cytosorb showed the best adsorption kinetics, while dialysis with polystyrene-based membranes offers the best biocompatibility because they do not show any unspecific adsorption of other plasma components.
背景/目的:细胞因子是需要有效去除的主要靶物质之一,以改善脓毒症、感染性休克和肝病患者的健康状况。尽管有各种商业上可用的医疗设备,但它们的临床应用效果有限。本体外研究的目的是比较 3 种不同的医疗设备在清除细胞因子白细胞介素-6(IL-6)、IL-8、IL-1β和肿瘤坏死因子-α方面的性能。测试的医疗设备是全血吸附剂 CytoSorb、高通量过滤器 EMiC2 和 HemofeelCH 1.8 血滤器。
方法
该研究在多滤机上用 1 L 人血浆进行 8 小时。在规定的时间点从血浆池采集用于细胞因子定量的样本。每个实验重复 3 次,并计算所有测试细胞因子的清除率。
结果
所有 3 种医疗设备均显示出良好的细胞因子清除效果。所有细胞因子的最高清除率是通过用 CytoSorb 进行血液灌流获得的。Hemofeel(连续静脉-静脉血液透析滤过)的 IL-8 和 IL-6 清除率高于 EMiC2(连续静脉-静脉血液透析),因为基于聚甲基丙烯酸甲酯(PMMA)的膜 Hemofeel 能够通过吸附去除这两种细胞因子。Cytosorb 的蛋白和白蛋白损失最高,EMiC2 的损失最低。
结论
血液净化清除细胞因子的机制包括对流、扩散和吸附。基于 PMMA 的过滤器能够结合所有 3 种机制来清除某些细胞因子。Cytosorb 显示出最佳的吸附动力学,而基于聚苯乙烯的膜的透析提供了最佳的生物相容性,因为它们不会对其他血浆成分产生任何非特异性吸附。