PLoS One. 2019 Sep 25;14(9):e0222920. doi: 10.1371/journal.pone.0222920. eCollection 2019.
The T-SPOT.TB can be read by an ELISPOT plate imager as an alternative to a labor-intensive and time-consuming manual reading, but its accuracy has not been sufficiently discussed to date.
1,423 test results obtained from manual reading using a microscope and an ELISPOT plate imager were compared. The agreement of qualitative test results was assessed using Cohen's kappa coefficient. The relationship of spot counts was studied using Bland-Altman analysis.
The overall percent agreement of the qualitative test results was 95.43% with a kappa coefficient of 0.91. Positive test results with the maximum net spot count of 8 and borderline test results showed relatively high discordance. The agreement of spot counts in panel A, panel B, and nil control was good, and variability did not increase with higher spot counts. On the basis of study findings, a novel strategy for interpreting the test results by an ELISPOT plate imager was proposed.
To increase diagnostic accuracy, positive test results with the maximum net spot count of 8 and borderline test results should be manually confirmed. Our strategy could be a practical guide for laboratories to build their own strategies for interpreting the test results by an ELISPOT plate imager.
T-SPOT.TB 可通过 ELISPOT 板成像仪进行读取,作为一种替代费力且耗时的手动阅读的方法,但迄今为止,其准确性尚未得到充分讨论。
将使用显微镜和 ELISPOT 板成像仪进行手动读取获得的 1423 项检测结果进行比较。使用 Cohen's kappa 系数评估定性检测结果的一致性。使用 Bland-Altman 分析研究斑点计数的关系。
定性检测结果的总体百分比一致性为 95.43%,kappa 系数为 0.91。最大净斑点计数为 8 的阳性检测结果和边界线检测结果显示出较高的不匹配。A 面板、B 面板和无对照的斑点计数一致性良好,且随着斑点计数的增加,变异性并未增加。基于研究结果,提出了一种通过 ELISPOT 板成像仪解读检测结果的新策略。
为了提高诊断准确性,应手动确认最大净斑点计数为 8 的阳性检测结果和边界线检测结果。我们的策略可为实验室提供一种实用的指南,以制定自己的通过 ELISPOT 板成像仪解读检测结果的策略。