Abdom Radiol (NY). 2020 Oct;45(10):3361-3368. doi: 10.1007/s00261-019-02247-1.
To compare radiation dose and image quality for abdominal CTs performed on a spectral detector CT (SDCT) and a comparable single-energy conventional CT scanner for patients of different sizes.
Four semi-anthropomorphic phantoms were scanned on an SDCT (IQon, Philips Healthcare) and a comparable single-energy CT (iCT 256, Philips Healthcare) under matched scan parameters. Image noise and radiation dose were compared. For the HIPAA-compliant, IRB-approved retrospective cohort patient study, radiation dose was compared after adjusting for patient water equivalent diameter. Difference in subjective and objective image quality was assessed on a subset of 50 patients scanned on both scanners by two readers.
CTDIvol and noise from SDCT were higher than conventional CT for all phantoms, with a relative difference of 7.8% (range 5.3-14%) for radiation dose and average difference of 9.0% (range 5.5-11%) for noise. 718 SDCT and 937 conventional CT patients were included in the patient study. CTDIvol for SDCT patients tends to be lower for smaller patients (- 2%, 95% confidence interval (- 5%, - 0.2%) for 200 mm water equivalent diameter) and higher for larger patients compared to conventional CT (8%, (6%, 11%) for 400 mm). No difference was seen for subjective image quality, SNR, CNR, or image noise between the two scanners, except for higher image noise in the portal vein and higher signal in the aorta on SDCT.
Radiation dose for abdominal CT performed on SDCT is similar to the dose on a conventional CT for average size patients, lower for smaller patients, and slightly higher for larger patients. Image quality is similar between the two scanners.
比较在光谱探测器 CT(SDCT)和可比的单能量常规 CT 扫描仪上为不同体型的患者进行腹部 CT 检查的辐射剂量和图像质量。
在 SDCT(IQon,飞利浦医疗保健)和可比的单能量 CT(iCT 256,飞利浦医疗保健)上对四个半人体模型体模进行扫描,扫描参数匹配。比较图像噪声和辐射剂量。对于符合 HIPAA 标准、IRB 批准的回顾性队列患者研究,在调整患者水等效直径后比较辐射剂量。由两位读者对在两台扫描仪上扫描的 50 名患者的一部分进行主观和客观图像质量的差异评估。
SDCT 的 CTDIvol 和噪声对于所有体模均高于常规 CT,辐射剂量的相对差异为 7.8%(范围 5.3-14%),噪声的平均差异为 9.0%(范围 5.5-11%)。患者研究中包括 718 名 SDCT 患者和 937 名常规 CT 患者。对于较小的患者,SDCT 的 CTDIvol 趋于较低(对于 200mm 水等效直径,为 -2%(95%置信区间[-5%,-0.2%]),而对于较大的患者,SDCT 的 CTDIvol 高于常规 CT(对于 400mm,为 8%(6%,11%))。在两台扫描仪之间,除了 SDCT 门静脉的图像噪声较高和主动脉的信号较高外,主观图像质量、SNR、CNR 或图像噪声没有差异。
对于平均体型的患者,SDCT 上进行的腹部 CT 的辐射剂量与常规 CT 的剂量相似,对于较小的患者,剂量较低,对于较大的患者,剂量略高。两台扫描仪的图像质量相似。