The purpose of this study was to determine whether application of a proprietary image-processing algorithm would allow a reduction in the necessary administered activity for molecular breast imaging (MBI) examinations. Images from standard-dose MBI examinations (300 MBq Tc-sestamibi) of 50 subjects were analyzed. The images were acquired in dynamic mode and showed at least one breast lesion. Half-dose MBI examinations were simulated by summing one-half of the dynamic frames and were processed with the algorithm under study in both a default and a preferred filter mode. Two breast radiologists independently completed a set of two-alternative forced-choice tasks to compare lesion conspicuity on standard-dose images, half-dose images, and the algorithm-processed half-dose images in both modes. Relative to the standard-dose images, the half-dose images were preferred in 4, the default-filtered half-dose images in 50, and preferred-filtered half-dose images in 76 of 100 readings. Compared with standard-dose images, in terms of lesion conspicuity, the half-dose images were rated better in 2, equivalent in 6, and poorer in 92 of 100 readings. The default-filtered half-dose images were rated better, equivalent, or poorer in 13, 73, and 14 of 100 readings. The preferred-filtered half-dose images were rated as better, equivalent, or poorer in 55, 34, and 11 of 100 readings. Compared with that on standard-dose images, lesion conspicuity on images obtained with the algorithm and acquired at one-half the standard dose was equivalent or better without compromise of image quality. The algorithm can also be used to decrease imaging time with a resulting increase in patient comfort and throughput.
本研究旨在确定应用专有的图像处理算法是否可以减少分子乳腺成像(MBI)检查所需的放射性活度。对 50 例接受标准剂量(300MBq Tc- sestamibi)MBI 检查的图像进行了分析。这些图像以动态模式采集,至少显示一个乳腺病变。半剂量 MBI 检查通过对一半的动态帧进行求和来模拟,并使用正在研究的算法在默认和首选滤波器模式下进行处理。两位乳腺放射科医生独立完成了一组二选一的强制选择任务,以比较标准剂量图像、半剂量图像和两种模式下经算法处理的半剂量图像上的病变显影。与标准剂量图像相比,在 100 次阅读中,4 次更喜欢半剂量图像,50 次更喜欢默认滤波的半剂量图像,76 次更喜欢首选滤波的半剂量图像。与标准剂量图像相比,在 100 次阅读中,2 次认为半剂量图像的病变显影更好,6 次认为相等,92 次认为更差。在 100 次阅读中,13 次认为默认滤波的半剂量图像更好、相等或更差,73 次认为相等,14 次认为更差。在 100 次阅读中,55 次认为首选滤波的半剂量图像更好、相等或更差,34 次认为相等,11 次认为更差。与标准剂量图像相比,使用该算法获得的半剂量图像的病变显影在不影响图像质量的情况下,其效果相当或更好。该算法还可用于减少成像时间,从而提高患者舒适度和检查通量。