Latifoğlu Fatma, Esas Mustafa Yasin, Demirci Esra
Erciyes University, Engineering Faculty, Biomedical Engineering Department, 39039 Kayseri, Turkey.
Biomed Tech (Berl). 2020 Apr 28;65(2):149-164. doi: 10.1515/bmt-2019-0027.
A system based on objective data was developed in the diagnosis and follow-up of attention-deficit hyperactivity disorder (ADHD) in this study. First of all, an electronic circuit, with a two-channel instrumentation amplifier designed to detect eye movements in the horizontal and vertical directions via surface electrodes, was developed to obtain the electrooculogram (EOG) signals. In order to provide a controlled analysis of eye movements during the reception of the signal, an attention test with visual stimulus software was developed. Eight patients with ADHD and eight healthy subjects were asked to monitor the stimulus images on the screen in the reference directions of the test system while recording EOG signals. According to the results of the t-test, no significant difference was found (p=0.11) between the healthy group and the reference movement information, whereas a significant difference was found between patients and the reference motion information (p=0.049). According to these results, it was seen that the number of eye movements of healthy individuals was statistically significant. In addition, they were inconsistent with the reference movement information. The level of significance was found to be low in patients. In this study, a new method is presented to test and diagnose individuals who were attention deficit.
本研究开发了一种基于客观数据的系统,用于注意力缺陷多动障碍(ADHD)的诊断和随访。首先,开发了一种电子电路,该电路带有一个双通道仪表放大器,设计用于通过表面电极检测水平和垂直方向的眼球运动,以获取眼电图(EOG)信号。为了在接收信号期间对眼球运动进行可控分析,开发了一款带有视觉刺激软件的注意力测试。八名ADHD患者和八名健康受试者被要求在记录EOG信号的同时,按照测试系统的参考方向监测屏幕上的刺激图像。根据t检验结果,健康组与参考运动信息之间未发现显著差异(p = 0.11),而患者与参考运动信息之间存在显著差异(p = 0.049)。根据这些结果,可以看出健康个体的眼球运动次数具有统计学意义。此外,它们与参考运动信息不一致。患者的显著性水平较低。在本研究中,提出了一种新的方法来测试和诊断注意力缺陷个体。