回顾性分析多室质子治疗中心降低能量切换和房间切换时间对吞吐量效率的影响。

Retrospective analysis of reduced energy switching and room switching times on throughput efficiency of a multi-room proton therapy center.

机构信息

ProCure Proton Therapy Center, 103 Cedar Grove Lane, Somerset NJ 08540, USA.

Transeo Radiothearpy Solutions, 1 Ferry Building #255, San Francisco CA 94111, USA.

出版信息

Br J Radiol. 2020 Mar;93(1107):20190820. doi: 10.1259/bjr.20190820. Epub 2019 Dec 2.

Abstract

OBJECTIVE

To quantify how a control software upgrade changed beam delivery times and impacted efficiency and capacity of a multiroom proton therapy center.

METHODS

A four-room center treating approximately 90 patients/day, treating for approximately 7 years with optimized operations, underwent a software upgrade which reduced room and energy switching times from approximately 30 to 20 s and approximately 4 s to ~0.5 s, respectively. The center uses radio-frequency identification data to track patient treatments and has software which links this to beam delivery data extracted from the treatment log server. Two 4-month periods, with comparable patient volume, representing periods before and after the software change, were retrospectively analyzed.

RESULTS

A total of 16,168 and 17,102 fields were analyzed. For bilateral head and neck and prostate patients, the beam waiting time was reduced by nearly a factor of 3 and the beam delivery times were reduced by nearly a factor of 2.5. Room switching times were reduced more modestly. Gantry capacity has increased from approximately 30 patients to 40-45 patients in a 16-h daily operation.

CONCLUSIONS

Many proton centers are striving for increased efficiencies. We demonstrated that reductions in energy and room switching time can significantly increase center capacity. Greater potential for further gains would come from improvements in setup and imaging efficiency.

ADVANCES IN KNOWLEDGE

This paper provides detailed measured data on the effect on treatment times resulting from reducing energy and room switching times under controlled conditions. It helps validate the models of previous investigations to establish treatment capacity of a proton therapy center.

摘要

目的

量化控制软件升级如何改变束流输送时间,并影响多间室质子治疗中心的效率和容量。

方法

一个治疗约 90 名患者/天的四间治疗室中心,经过约 7 年的优化操作,进行了软件升级,将房间和能量切换时间分别从约 30 秒和 4 秒减少到约 20 秒和 0.5 秒左右。该中心使用射频识别数据来跟踪患者的治疗情况,并使用软件将其与从治疗日志服务器中提取的束流输送数据相关联。回顾性分析了两个具有可比性的 4 个月患者量,代表软件更改前后的时间段。

结果

共分析了 16168 和 17102 个野。对于双侧头颈部和前列腺患者,束流等待时间减少了近 3 倍,束流输送时间减少了近 2.5 倍。房间切换时间的减少幅度较小。在 16 小时的日常操作中,龙门架容量从大约 30 名患者增加到 40-45 名患者。

结论

许多质子中心都在努力提高效率。我们证明,减少能量和房间切换时间可以显著增加中心容量。通过改进设置和成像效率,可以进一步提高效率。

知识进展

本文提供了在受控条件下减少能量和房间切换时间对治疗时间影响的详细测量数据。它有助于验证以前调查的模型,以建立质子治疗中心的治疗能力。

相似文献

4
Modelling the throughput capacity of a single-accelerator multitreatment room proton therapy centre.
Br J Radiol. 2012 Dec;85(1020):e1263-72. doi: 10.1259/bjr/27428078.
5
A predictive algorithm for spot position corrections after fast energy switching in proton pencil beam scanning.
Med Phys. 2018 Nov;45(11):4806-4815. doi: 10.1002/mp.13217. Epub 2018 Oct 16.
6
Single-Room Proton Radiation Therapy Systems: No Small Change.
Int J Radiat Oncol Biol Phys. 2016 May 1;95(1):147-148. doi: 10.1016/j.ijrobp.2015.09.042.
7
Downstream Effect of a Proton Treatment Center on an Academic Medical Center.
Int J Radiat Oncol Biol Phys. 2019 Jul 15;104(4):756-764. doi: 10.1016/j.ijrobp.2019.03.016. Epub 2019 Mar 15.
8
A comprehensive Monte Carlo study of out-of-field secondary neutron spectra in a scanned-beam proton therapy gantry room.
Z Med Phys. 2021 May;31(2):215-228. doi: 10.1016/j.zemedi.2021.01.001. Epub 2021 Feb 20.
9
Proton Therapy Facility Planning From a Clinical and Operational Model.
Technol Cancer Res Treat. 2015 Oct;14(5):635-41. doi: 10.7785/tcrt.2012.500444. Epub 2014 Nov 26.

引用本文的文献

2
Proton therapy special feature: introductory editorial.
Br J Radiol. 2020 Mar;93(1107):20209004. doi: 10.1259/bjr.20209004.

本文引用的文献

1
Highly efficient and sensitive patient-specific quality assurance for spot-scanned proton therapy.
PLoS One. 2019 Feb 14;14(2):e0212412. doi: 10.1371/journal.pone.0212412. eCollection 2019.
2
Technical Note: An efficient daily QA procedure for proton pencil beam scanning.
Med Phys. 2018 Mar;45(3):1040-1049. doi: 10.1002/mp.12787. Epub 2018 Feb 19.
3
Three ways to make proton therapy affordable.
Nature. 2017 Sep 25;549(7673):451-453. doi: 10.1038/549451a.
5
A comprehensive and efficient daily quality assurance for PBS proton therapy.
Phys Med Biol. 2017 Mar 7;62(5):1661-1675. doi: 10.1088/1361-6560/aa5131. Epub 2017 Feb 6.
7
A novel daily QA system for proton therapy.
J Appl Clin Med Phys. 2013 Mar 4;14(2):4058. doi: 10.1120/jacmp.v14i2.4058.
9
Modelling the throughput capacity of a single-accelerator multitreatment room proton therapy centre.
Br J Radiol. 2012 Dec;85(1020):e1263-72. doi: 10.1259/bjr/27428078.
10
Proton facility economics: the importance of "simple" treatments.
J Am Coll Radiol. 2012 Aug;9(8):560-3. doi: 10.1016/j.jacr.2012.03.014.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验