单克隆抗体电荷变异体与 VEGF-A 及其 ELISA 试剂的相互作用是否会影响其定量?

Does interaction of monoclonal antibody charge variants with VEGF-A and ELISA reagents affect its quantification?

机构信息

Department of Chemical Engineering, Indian Institute of Technology Delhi, Hauz Khas 110016, India.

Department of Chemical Engineering, Indian Institute of Technology Delhi, Hauz Khas 110016, India.

出版信息

Anal Biochem. 2020 Feb 1;590:113513. doi: 10.1016/j.ab.2019.113513. Epub 2019 Nov 27.

Abstract

The ability of an enzyme linked immunosorbent assay (ELISA) to measure the concentration of charge variants of a monoclonal antibody (mAb) biotherapeutic has been investigated. Percentage recovery was found to be in the range of 80%-120% with inter and intra assay variation between 5% and 15%. Linear regression analysis of ELISA output at different dilutions yielded a good fit (R > 0.98). Assay output was not hindered by the presence of serum proteins and other analytes such as GCSF, demonstrating acceptable precision and specificity. It was concluded that the charge variants of the mAb can be accurately quantified by ELISA.

摘要

已经研究了酶联免疫吸附测定(ELISA)测量单克隆抗体(mAb)治疗药物的电荷变异体浓度的能力。发现回收率在 80%-120%范围内,并且在 5%-15%之间存在分析内和分析间的变化。在不同稀释度下进行 ELISA 输出的线性回归分析得出了很好的拟合度(R>0.98)。该测定不受血清蛋白和其他分析物(如 GCSF)的存在的影响,显示出可接受的精密度和特异性。结论是 ELISA 可以准确地定量 mAb 的电荷变异体。