Reproducibility of the Early Cataract Detector (Kowa ECD 2000).

作者信息

Datiles M, Podgor M, Edwards P

机构信息

Clinical Branch, National Eye Institute, Bethesda, Maryland 20892.

出版信息

Ophthalmic Surg. 1988 Sep;19(9):664-6.

PMID:3186178
Abstract

We conducted a reproducibility study on the Kowa Early Cataract Detector, a laser light scattering spectroscope that determines the diameter of particles within the lens and designed to quantify changes in particle size in early cataracts. We found good reproducibility in the nuclear region (where changes are homogeneous), but found poor reproducibility in the cortical region (where changes are heterogeneous). We advise careful evaluation of other similar devices before extensive use.