Sato Toshimitsu, Kikuchi Yuho, Nakamura Masataka, Saito Nobutoshi, Arai Yuka
Department of Radiology, Yamagata University Hospital.
Health Sciences, Tohoku University Graduate School of Medicine.
Nihon Hoshasen Gijutsu Gakkai Zasshi. 2019;75(12):1403-1410. doi: 10.6009/jjrt.2019_JSRT_75.12.1403.
This research measured the radiation exposure of the computed tomography(CT) localizer radiograph of the trunk of the body. The entrance surface dose for CT localizer radiograph was measured using radiophotoluminescent glass dosimeter(RPLD) on four points of measurement, including the center of the phantom, on the surface of a phantom placed in the center of a CT bed, assuming that the subject has a thickness of 20 cm. The entrance surface dose of the localizer radiograph under the chest CT protocol manufacturer's initial setting conditions of 120 kV 35 mA was 0.80 mGy at the center and 0.53 for the 4-location average for the upper X-ray tube (excluding the CT bed), and 0.74 mGy at the center and 0.48 mGy for the 4-location average for the lower X-ray tube (including the CT bed). Compared to the Japan DRLs 2015 chest X-ray (P→A), the entrance surface dose was 2.67 times at the center and 1.77 times for the 4-location average for the upper X-ray tube and 2.47 times at the center and 1.60 times for the 4-location average for the lower X-ray tube. The CT radiation dose also cannot be ignored for the localizer radiograph entrance surface dose.
本研究测量了身体躯干计算机断层扫描(CT)定位X光片的辐射剂量。假设受试者厚度为20厘米,在置于CT检查床中心的体模表面的四个测量点(包括体模中心),使用放射光致发光玻璃剂量计(RPLD)测量CT定位X光片的体表入射剂量。在胸部CT协议制造商初始设置条件120 kV 35 mA下,上X射线管(不包括CT检查床)的定位X光片体表入射剂量在中心为0.80 mGy,四点平均为0.53 mGy;下X射线管(包括CT检查床)的定位X光片体表入射剂量在中心为0.74 mGy,四点平均为0.48 mGy。与2015年日本胸部X光片剂量参考水平(P→A)相比,上X射线管的体表入射剂量在中心为2.67倍,四点平均为1.77倍;下X射线管的体表入射剂量在中心为2.47倍,四点平均为1.60倍。CT定位X光片的体表入射剂量的CT辐射剂量也不容忽视。