Accuracy of lens power calculations with the biconvex and meniscus intraocular lenses.

作者信息

Shammas H J

机构信息

Department of Ophthalmology, University of Southern California, Los Angeles.

出版信息

Am J Ophthalmol. 1988 Nov 15;106(5):613-5. doi: 10.1016/0002-9394(88)90596-x.

Abstract

In a prospective study, intraocular lens power calculations were within +/- 0.50 diopter of the final refraction in 77 of 88 eyes (87.5%) with a biconvex lens and in 41 of 90 eyes (45.5%) with a meniscus lens. The more precise power calculations with the biconvex intraocular lens seemed to be related to the superior optical design of the lens, with its close configuration to a normal crystalline lens.

摘要

在一项前瞻性研究中,88只植入双凸透镜的眼睛中有77只(87.5%)的人工晶状体屈光力计算值与最终验光结果相差在±0.50屈光度以内,而90只植入新月形透镜的眼睛中有41只(45.5%)的计算值在该范围内。双凸人工晶状体更精确的屈光力计算似乎与其优越的光学设计有关,它与正常晶状体的结构相近。