Calibration of Stereo Radiography System for Radiostereometric Analysis Application.

作者信息

Allab A, Vazquez C, Cresson T, Guise J de

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2019 Jul;2019:4859-4862. doi: 10.1109/EMBC.2019.8857531.

Abstract

This paper describes a new alternative to conventional radiography system currently used for radiostereometric analysis studies. Instead of using two non-calibrated X-ray sources with a cumbersome calibration cage, we propose to use the biplanar radiography EOS system. Its fixed configuration provides a preliminary calibration and a much simpler acquisition protocol. A flexible and accurate calibration method is presented to optimize EOS default calibration using a simple object and a self-calibration method. To validate our system, we calculate the 3D reconstruction error of a known object. Results showed an accuracy of 70±11μm and 0.05±0.02° for translation and rotation respectively, and an average epipolar error of 23±03μm.

摘要

本文介绍了一种新型的替代方案,用于目前放射立体分析研究中使用的传统放射成像系统。我们提议使用双平面放射成像EOS系统,而不是使用两个未经校准的X射线源以及笨重的校准笼。其固定配置提供了初步校准和更为简单的采集协议。提出了一种灵活且精确的校准方法,通过使用一个简单物体和自校准方法来优化EOS默认校准。为验证我们的系统,我们计算了一个已知物体的三维重建误差。结果表明,平移和旋转的精度分别为70±11μm和0.05±0.02°,平均极线误差为23±03μm。