Oku T, Harasaki H, Smith W, Nosé Y
Cleveland Clinic Foundation, Ohio.
ASAIO Trans. 1988 Jul-Sep;34(3):500-4.
Hemolysis is one of the important performance parameters of blood pumps. However, comparisons of published data are compromised by the different test methods used. This article reports the results obtained when testing four nonpulsatile blood pumps (Roller pump, Biopump Model 600, Biopump BP-80, Hemadyne pump) in the identical test circuit under the same hemodynamic conditions. A closed test circuit was composed of a flexible bag reservoir and two short tubes for connection to the pump head. The priming volume was 1 L, and the total surface area was 950 cm2. Fresh blood from a fasted calf was directly primed into the circuit and maintained at 37 degrees C. It was pumped at 6 L/min for 12 hr against a 100 mmHg afterload. The hemolysis index (HI), defined as grams of hemoglobin released per 100 L pumped, was calculated from a curve fit to the hourly data. Each pump was tested five times. The mean HI obtained ranged from 0.0467 for the roller pump to 0.0076 for the Biopump BP-80. There was no statistically significant difference among the three centrifugal pumps tested.
溶血是血泵的重要性能参数之一。然而,已发表数据的比较因所使用的不同测试方法而受到影响。本文报告了在相同的血流动力学条件下,在相同的测试回路中对四种非搏动性血泵(滚压泵、600型生物泵、BP - 80型生物泵、Hemadyne泵)进行测试时所获得的结果。一个封闭的测试回路由一个柔性袋式储液器和两根连接到泵头的短管组成。预充量为1升,总表面积为950平方厘米。来自禁食小牛的新鲜血液直接预充到回路中,并保持在37摄氏度。在100 mmHg的后负荷下,以6升/分钟的速度泵血12小时。溶血指数(HI)定义为每泵出100升所释放的血红蛋白克数,通过对每小时数据进行曲线拟合来计算。每个泵测试五次。所获得的平均HI范围从滚压泵的0.0467到BP - 80型生物泵的0.0076。在所测试的三种离心泵之间没有统计学上的显著差异。